Comprehensive calibration
The comprehensive calibration procedure calibrates the DCV, DCI, ACV, ACI, and ohms
functions of the Model 2790. You can also choose to calibrate only the DCV/DCI and
resistance or ACV/ACI functions.
These procedures are usually the only calibration required in the field. Manufacturing
calibration is normally done only at the factory, but it should also be done in the field if the
unit has been repaired. See “Manufacturing calibration” later in this section for more
information.
Calibration cycle
Perform Model 2790 comprehensive calibration at least once a year, or every 90 days to
ensure that the unit meets the corresponding specifications.
Recommended equipment
lists the recommended equipment you need for comprehensive, DC-only, and
AC-only calibration procedures. You can use alternate equipment, such as a DC transfer
standard and characterized resistors, as long as that equipment has specifications equal to
or better than those listed in the table. In general, equipment uncertainty should be at least
four times better than corresponding Model 2790 specifications.
Table 17-1
Recommended equipment for comprehensive calibration
Fluke 5700A Calibrator:
DC voltage
AC voltage
(1kHz, 50kHz)*
DC current
AC current
(1kHz)
Resistance
10V:±5ppm
100V: ±7ppm
10mV:±710ppm
100mV:±200ppm
1V:±82ppm
10V:±82ppm
100V:±90ppm
700V:±85ppm
10mA:±60ppm
100mA:±70ppm
1A:±110ppm
100mA:±190ppm
1A:±690ppm
2A:±670ppm
1k
Ω
:±13ppm
10k
Ω
:±11ppm
100k
Ω
:±13ppm
1M
Ω
:±18ppm
Miscellaneous equipment:
Keithley 8610 low-thermal shorting plug
Double banana plug to double banana plug shielded cable
BNC to double banana plug shielded cable
*1kHz specifications. 10mV and 700V points require 1kHz only. All calibrator specifications are 90-day, 23°C ±5°C specifications and
indicate total absolute uncertainty at specified output.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...