Model 2750 Multimeter/Switch System User’s Manual
Getting Started
1-17
Remote programming
The following command controls keyclick:
SYSTem:KCLick <b>
' Enable or disable keyclick.
where: <b> = ON or OFF
NOTE
Keyclick ON is the FACTORY, *RST, and SYSTem:PRESet default.
Display
Readings are displayed in engineering units (i.e., 100.23mV), while annunciators indicate
various states of operation. See
“Front panel summary,” page 1-9
, for a complete listing of
display annunciators.
NOTE
The display test allows you to test display digit segments, annunciators and the
red slot indicator LEDs. The key test checks the functionality of front panel keys.
These tests are accessed by pressing SHIFT and then TEST. Refer to the Model
2750 Service Manual for details.
Status and error messages
Status and error messages are displayed momentarily. During operation and programming,
you will encounter a number of front panel messages. Typical messages are either of status
or error variety, as listed in
Remote programming — display
Using remote programming, the Model 2750 can display a custom ASCII message (up to
12 characters). Also, the front panel display and controls can be disabled.
Display commands
The commands are listed in
. Details on these commands follow the table.
NOTE
Optional command words and queries are not included in
provides an unabridged list of all display commands.
Table 1-3
Display commands
Command
Description
Default*
DISPlay:TEXT:DATA <a>
Define message (<a> = ASCII characters,
up to 12).
(none)
DISPlay:TEXT:STATe <b> Enable or disable message mode (<b> = ON
or OFF).
OFF
DISPlay:ENABle <b>
Enable or disable the front panel display
(<b> = ON or OFF).
OFF
*SYSTem:PRESet and *RST have no effect on DISPlay commands. The listed defaults are power-on defaults.
2750-900-01.book Page 17 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...