Model 2750 Multimeter/Switch System Service Manual
Routine Maintenance
3-5
Routin
e Mainten
ance
Replacing Model 7700 plug-in module amps fuses
WARNING
The information in this section is intended only for qualified service per-
sonnel. Do not perform these procedures unless you are qualified to do so.
WARNING
Make sure that all plug-in module connections are de-energized and dis-
connected before replacing module amps fuses.
1.
Turn off the power, and disconnect the power line and external connections from
the Model 7700.
2.
Open the Model 7700 top cover.
3.
Locate the amps fuses for CH21 and CH22 (
).
4.
Remove the circuit board from the bottom plastic housing by removing the two
bottom screws.
Figure 3-3
Model 7700 amps fuses
H L H L
AMPS
H L H L
H L H L
H L H L
LO
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
H L H L H L H L
CH17 CH18 CH19 CH20
SENSE
(OHMS, 4 WIRE)
INPUT
(V, 2 WIRE)
H L H L H L H L
CH7
CH8 CH9 CH10
CH22 Fuse
CH21 Fuse
Model 7700
H L H L H L H L
H L H L H L H L
INPUT SENSE CH1
CH2
CH3
CH4
CH5
CH6
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......