Model 2750 Multimeter/Switch System Service Manual
Performance Verification
1-15
275
0 V
eri
fication
3.
Select the Model 2750 4-wire resistance function by pressing the
4 key, then
choose the SLOW integration rate with the RATE key.
4.
Set the Model 2750 for the 1
range, and make sure the FILTER is on. Enable
OCOMP (offset-compensated ohms) by pressing SHIFT then OCOMP. (Use
OCOMP for 1
, 10
, and 100
range verification.)
5.
Recalculate reading limits based on actual calibrator resistance values.
6.
Source the nominal full-scale resistance values for the 1
-10M
ranges summa-
, and verify that the readings are within calculated limits.
7.
Connect the Model 2750 INPUT and SENSE jacks to the calibrator as shown in
.
8.
Disable external sense on the calibrator.
9.
Set the Model 2750 for the 100M
range.
10.
Source a nominal 100M
resistance value and verify that the reading is within cal-
culated limits for the 100M
range.
Figure 1-6
Connections for Model 2750 resistance verification (100M
range)
Input
HI
Input
LO
Output
LO
Output
HI
Calibrator (Output 2-wire Resistance)
Note: Use shielded cables to minimize noise.
Disable calibrator external sense mode.
Sense
LO
Sense
HI
!
F
500V
PEAK
FRONT/REAR
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUT
350V
PEAK
1000V
PEAK
R
CAT I
Model 2750
Front Panel
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......