![Keithley 2750 Скачать руководство пользователя страница 102](http://html1.mh-extra.com/html/keithley/2750/2750_service-manual_3957779102.webp)
4-16
Troubleshooting
Model 2750 Multimeter/Switch System Service Manual
Analog signal switching states
provide switching states of the various relays, FETs, and
analog switches for the basic measurement functions and ranges. These tables can be used
to assist in tracing an analog signal from the input to the A/D multiplexer.
Table 4-5
DCV signal switching
Range
Q101
Q102
Q151
Q150
Q114
Q136
Q109
100mV
1V
10V
100V
1000V
ON
ON
ON
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
OFF
OFF
Range
K101*
Q113
Q104
Q108
Q106
Q148
100mV
1V
10V
100V
1000V
SET
SET
SET
SET
SET
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
OFF
OFF
OFF
OFF
OFF
ON
ON
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
ON
*K101 set states:
Pin 8 switched to Pin 7
Pin 3 switched to Pin 4
K101 reset states: Pin 8 switched to Pin 9
Pin 3 switched to Pin 2
Содержание 2750
Страница 13: ...1 PerformanceVerification...
Страница 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Страница 49: ...2 Calibration...
Страница 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Страница 75: ...Routine Maintenance 3 RoutineMaintenance...
Страница 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Страница 87: ...Troubleshooting 4 Troubleshooting...
Страница 111: ...Disassembly 5 Disassembly...
Страница 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Страница 119: ...Replaceable Parts 6 ReplaceableParts...
Страница 135: ...Specifications A AccuracyandOptimization...
Страница 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Страница 141: ...Calibration Reference B CalibrationReference...
Страница 167: ...Calibration Program C CalibrationProgram...
Страница 178: ......
Страница 179: ......