3-2
Basic DMM Operation
Model 2701 User’s Manual
DMM measurement capabilities
NOTE
Accuracy specifications for all measurement functions and the Model 7700
switching module are provided in Appendix A.
NOTE
For remote operations, you can use the internal web page of the Model 2701 to
send commands, queries, and take readings. See “Internal web page” in
Section 10 for details.
WARNING
The Model 2701 measurement inputs are rated measurement Category I.
The equipment cannot be connected to circuits with transients greater than
4kV peak. Do not connect measurement inputs to measurement Category
II, III, or IV circuits.
The DMM of the Model 2701 can make the following measurements:
DCV — DC voltage measurements from 0.1µV to 1000V.
ACV — AC voltage measurements from 0.1µV to 750V.
DCI — DC current measurements from 10nA to 3A.
ACI — AC current measurements from 1µA to 3A.
Ω
2 — 2-wire resistance measurements from 100µ
Ω
to 120M
Ω
.
Ω
4 — 4-wire resistance measurements from 100µ
Ω
to 120M
Ω
.
FREQ — Frequency measurements from 3Hz to 500kHz.
PERIOD — Period measurements from 333ms to 2µs.
TEMP — Temperature measurements from -200°C to 1820°C.
CONT — Continuity testing using the 1k
Ω
range.
CAUTION
When using a switching module, do not exceed the maximum signal levels
of the module. To prevent damage to the Model 7700 switching module,
do not exceed these maximum signal levels:
Channels 1-20:
300VDC or 300V RMS (425V peak) for AC
waveforms, 1A switched, 60W, 125VA
Channels 21, 22:
60VDC or 30V RMS, 3A switched, 60W, 125VA
For the other switching modules, the maximum signal levels are included
with their specifications.
NOTE
This section shows DUT connections to the front panel inputs of the Model 2701
and to the Model 7700 switching module. Details on Model 7700 connections
are provided in Appendix B.
2701-900-01.book Page 2 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...