Optimizing measurement accuracy
The configurations listed below assume that the multimeter has had factory setups
restored.
NOTE
For maximum system performance, it is recommended that all measurement
cables be limited to less than 3 meters.
DC voltage, DC current, and resistance:
•
Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range.
•
Use REL on DC voltage and 2-wire resistance measurements.
•
Use 4-wire resistance measurements for best accuracy.
AC voltage and AC current:
•
Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range.
Temperature:
•
Select 6½ digits, 10 PLC, filter ON (up to 100 readings).
Optimizing measurement speed
The configurations listed below assume that the multimeter has had factory setups
restored.
DC voltage, DC current, and resistance:
•
Select 3½ digits, 0.002 PLC, filter OFF, fixed range.
AC voltage and AC current:
•
Select 3½ digits, 0.002 PLC, filter OFF, fixed range.
Temperature:
•
Select 3½ digits, 0.002 PLC, filter OFF.
For all functions, turn off the display, autozero, and set the trigger delay to zero. Use the
:SAMPle:COUNt and READ? bus commands.
2701-900-01.book Page 10 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...