Model 2701 User’s Manual
Triggering
8-19
Output Trigger
— The Model 2701 will send one or more output triggers. The output
trigger is applied to the Trigger Link connector on the rear panel. It can be used to trigger
an external instrument to perform an operation.
The trigger model can be configured to output a trigger after the completion of a series of
measurements or after every measurement. For example, with the sample counter set to 10
and the trigger counter set to one, a trigger will be sent after the 10 measurements are
performed. If instead, the trigger counter is set to 10 and the sample counter is set to 1, a
trigger will be sent after each measurement.
Triggering commands
Commands for triggering are summarized in Table 8-2.
Table 8-2
SCPI commands — triggering
Commands
Description
Default Ref
ABORt
Reset trigger system.
a
INITiate[:IMMediate]
Initiate one trigger cycle.
b
INITiate:CONTinuous <b>
Enable/disable continuous initiation; ON or OFF.
Note 1
c
FETCh?
Request the last reading(s).
d
READ?
Perform an ABORt, INITiate, and a FETCh?
d
TRIGger:SOURce <name>
Select control source; IMMediate, TIMer, MANual,
BUS, or EXTernal.
IMM
e
TRIGger:TIMer <n>
Set timer interval; 0 to 999999.999 (sec).
0.1
TRIGger:COUNt <NRf>
Set trigger count; 1 to 450000 or INFinity.
Note 2
TRIGger:DELay <n>
Set delay; 0 to 999999.999 (sec).
0
TRIGger:DELay:AUTO <b>
Enable or disable auto delay.
f
TRIGger:SIGNal
Loop around control source.
g
SAMPle:COUNt <NRf>
Set sample count; 1 to 450000.
1
h
[:SENSe[1]
Optional root command for HOLD commands.
i
:HOLD:WINDow <NRf>
Set Hold window in %; <NRf> = 0.01 to 20.
1
:HOLD:COUNt <NRf>
Set Hold count; <NRf> = 2 to 100.
5
:HOLD:STATe
<b>
Enable or disable Hold.
OFF
:SYSTem:BEEPer:STATe <b> Enable or disable the beeper.
ON
*RST
Restore *RST defaults (see “Default” column of this
table). Places 2701 in the idle state.
Notes:
1. SYSTem:PRESet enables continuous initiation.*RST disables continuous initiation.
2. SYSTem:PRESet sets the trigger count to INF (infinity). *RST sets the count to 1.
2701-900-01.book Page 19 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...