Model 2701 Instrument Networking Instruction Manual
Configuring TCP/IP addresses
3-7
Changing Model 2701 TCP/IP addresses
The Model 2701 leaves the factory with the default settings in
.
If the instrument is not recognized by the network, check the TCP/IP settings and make any
changes required. Each Model 2701 ships with a unique MAC address—see the Glossary for a
detailed definition.
Front panel
Check the TCP/IP address
1.
Press Shift.
2.
Press Ethernet (VIEW is displayed).
3.
Press Enter to select VIEW.
4.
Use the
¹
and
ƒ
range keys to scroll between:
•
ADDRESS
•
SUBNET
•
GATEWAY
•
MAC (read-only, cannot change)
5.
Press enter to view the desired setting.
Set the IP address
1.
Press Shift.
2.
Press Ethernet (VIEW is displayed).
3.
Use the
¹
and
ƒ
range keys to scroll the display to SET.
4.
Press Enter to select SET.
5.
Modify the settings:
•
Use the
¹
and
ƒ
range keys to change values. Move the cursor with the
¹
and
ƒ
arrow
keys.
•
Press enter to set the desired field and open the next setting.
Table 3-4
Model 2701 default TCP/IP addresses
Field
Default value
IP ADDRESS
192.168.000.002
GATEWAY
192.168.000.001
SUBNET
255.255.255.000
DHCP
OFF