Model 2700 Multimeter/Switch System User’s Manual
Status Structure
11-23
On power-up, all error messages are enabled and will go into the Error Queue as they
occur. Status messages are not enabled and will not go into the queue. As listed in
, there are commands to enable and/or disable messages. For these commands,
the <list> parameter is used to specify which messages to enable or disable. The messages
are specified by their codes. The following examples show various forms for using the
<list> parameter.
<list>
= (-110)
Single message
= (-110:-222)
Range of messages (-110 through -222)
= (-110:-222, -220)
Range entry and single entry (separated by a comma)
When you enable messages, messages not specified in the list are disabled. When you
disable messages, each listed message is removed from the enabled list.
NOTE
To prevent all messages from entering the Error Queue, send the enable
command along with the null list parameter as follows:
STATus:QUEue:ENABle ().
Programming example — read error queue
NOTE
The following example can be run from the KE2700 Instrument Driver using the
example named “ReadErrorQueue” in
The following command reads the error queue:
STAT:QUE?
Table 11-6
Error queue commands
Command
Description
Notes
STATus:QUEue:[:NEXT]?
Read and clear oldest error/status (code and message).
1
STATus:QUEue:ENABle <list>
Specify error and status messages for Error Queue.
2
STATus:QUEue:ENABle?
Read the enabled messages.
STATus:QUEue:DISable <list>
Specify messages not to be placed in queue.
2
STATus:QUEue:DISable?
Read the disabled messages.
STATus:QUEue:CLEar
Clear messages from Error Queue.
SYSTem:ERRor?
Read Error Queue.
1
SYSTem:CLEar
Clear messages from Error Queue.
Notes:
1. Power-up and *CLS empties the Error Queue. STATus:PRESet has no effect.
2. Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no effect.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......