4-18
Range, Digits, Rate, Bandwidth, and Filter
Model 2700 Multimeter/Switch System User’s Manual
Filter control and configuration
The FILTER key toggles the state of the Filter. When the Filter is enabled, the FILT
annunciator is on. The FILT annunciator will flash when the filter is not settled. When
disabled, the FILT annunciator is off. The filter can be configured while it is enabled or
disabled.
The filter is configured from the filter configuration menu (
). Perform the
following steps to configure the filter:
1.
Select the desired function.
2.
Press SHIFT and then TYPE. The present WINDOW setting will be displayed.
3.
Use the RANGE
Δ
or
∇
key to display the desired window setting (0.01%, 0.1%,
1%, 10%, or NONE), and press ENTER.
4.
Use the
,
,
Δ
, and
∇
keys to display the number of readings to filter (1 to
100), and press ENTER.
5.
Use the
Δ
or
∇
key to display the desired filter type (moving or repeating), and
press ENTER. The filter turns on and the instrument returns to the normal
measurement state.
NOTE
While the filter is enabled (FILT annunciator on), changes to the configuration
take effect as soon as they are made. With filter disabled (FILT annunciator off),
changes to the configuration take place when the filter is enabled.
While the filtering operation is in progress, the FILT annunciator blinks.
Readings will continue to be processed (i.e., displayed, stored, sent over the
bus), but they could be questionable. When the FILT annunciator stops blinking,
the filter has settled.
Changing function or range causes the filter to reset. The filter then assumes the
state (enabled or disabled) and configuration for that function or range.
Figure 4-4
Filter configuration flow chart
SHIFT
TYPE
WINDOW
0.01%
0.1%
1%
NONE
RDGS
001 to 100
TYPE
REPEAT
MOVNG AV
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......