4-14
Range, Digits, Rate, Bandwidth, and Filter
Model 2700 Multimeter/Switch System User’s Manual
Filter type
There are two digital filter types: moving and repeating. The moving average filter uses a
first-in first-out stack, where the newest reading conversion replaces the oldest. An aver-
age of the stacked reading conversions yields a filtered reading. After the specified number
of reading conversions (“
”) fill the stack, the moving filter gives a new reading
for every new conversion. This process is depicted in
A.
The repeating filter takes a specified number of conversions, averages them and yields a
filtered reading. It then flushes its stack and starts over. This character is useful when scan-
ning (readings for other channels are not averaged with the present channel). The stack is
then cleared and the process starts over (see
NOTE
The moving filter cannot be used when scanning. If a scan channel is set up to
use the moving filter, the filter will not turn on. Scanning is covered in
.
Filter count
The filter count specifies how many consecutive A/D conversions (within the filter
window) to place in the memory stack. When the stack is full, the A/D conversions are
averaged to calculate the final filtered reading. The filter count can be set from 1 to 100.
Note that with a filter count of 1, no averaging is done. However, only readings within the
“
” will be displayed, stored, or transmitted.
NOTE
While the filter processes readings, the FILT annunciator blinks. Readings that
are being displayed while the FILT annunciator blinks are not final filtered
readings. When the FILT annunciator stops blinking, the filter has settled.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......