Connect the instrument 

Important test system safety information 

This product is sold as a stand-alone instrument that may become 
part of a system that could contain hazardous voltages and energy 

sources. It is the responsibility of the test system designer, integrator, 

installer, maintenance personnel, and service personnel to make sure 
the system is safe during use and is operating properly. 

 You must also realize that in many test systems a single fault, such as 

a software error, may output hazardous signal levels even when the 
system indicates that there is no hazard present. 

It is important that you consider the following factors in your system 
design and use: 

 

The international safety standard IEC 61010-1 defines voltages as 

hazardous if they exceed 30 V RMS and 42.4 V peak, or 60 V DC 

for equipment rated for dry locations. Keithley Instruments. 
products are only rated for dry locations. 

 

Read and comply with the specifications of all instruments in the 

system. The overall allowed signal levels may be constrained by 
the lowest rated instrument in the system. For example, if you are 

using a 500 V power supply with a 300 V DC rated switch, the 

maximum allowed voltage in the system is 300 V DC. 

 

Make sure any test fixture connected to the system protects the 

operator from contact with hazardous voltages, hot surfaces, and 
sharp objects. Use shields, barriers, insulation, and safety 

interlocks to accomplish this. 

 

Cover the device under test (DUT) to protect the operator from 

flying debris in the event of a system or DUT failure. 

 

Double-insulate all electrical connections that an operator can 
touch. Double insulation ensures the operator is still protected 

even if one insulation layer fails. Refer to IEC 61010-1 for 

specific requirements. 

 

Make sure all connections are behind a locked cabinet door or 
other barrier. This protects the system operator from accidentally 

removing a connection by hand and exposing hazardous voltages. 

Содержание 2657A

Страница 1: ...Model 2657A High Power System SourceMeter Instrument Quick Start Guide...

Страница 2: ...work on live circuits perform safe installations and repair products Only properly trained service personnel may perform installation and service procedures Keithley Instruments products are designed...

Страница 3: ...in the specifications and operating information and as shown on the instrument or test fixture panels or switching card When fuses are used in a product replace with the same type and rating for cont...

Страница 4: ...not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument Products that consist of a circuit board with no case or chassis e g a data acquisition board for inst...

Страница 5: ...ion about the instrument s features and programming commands The User s and Reference manuals are in PDF format on the CD ROM that is included with the instrument If you do not have Adobe Reader to vi...

Страница 6: ...Unpack and inspect the instrument To unpack and inspect the instrument 1 Inspect the box for damage 2 Open the top of the box...

Страница 7: ...emove the packaging insert 4 Carefully lift the Model 2657A instrument out of the box 4 Inspect the instrument for any obvious signs of physical damage Report any damage to the shipping agent immediat...

Страница 8: ...low ground cable CA 568 120 4 RJ45 LAN crossover cable 5 Interlock DB 25 male connector kit hardware 6 Model 2657A Quick Start Guide not shown this document 7 Model 2657A Product Information and Test...

Страница 9: ...only rated for dry locations Read and comply with the specifications of all instruments in the system The overall allowed signal levels may be constrained by the lowest rated instrument in the system...

Страница 10: ...eep users safe always read and follow all safety warnings provided with each of the instruments in your system Install the instrument The Model 2657A can be used on a bench or in a rack Please see the...

Страница 11: ...applied is 4 V The absolute maximum input is 0 4 V to 6 0 V The interlock is intended for use through a normally open switch This switch may be installed on the lid of a test fixture the enclosure of...

Страница 12: ...rth safety ground Nonconductive test fixtures must be rated to double the maximum voltage in the test circuit Failure to attach the ground wires to a known protective earth may result in electric shoc...

Страница 13: ...hassis is connected to power line ground through the ground wire in the power cord Failure to use a grounded outlet may result in personal injury or death due to electric shock Operating the instrumen...

Страница 14: ...nce When the instrument is turned on you should see Five dots All segments of the display light A brief display showing KEITHLEY Model 2657A Other startup checks The entire power up process takes appr...

Страница 15: ...d You do not need to connect a device under test DUT for this test Step 1 Set source function range and level 1 Press the SRC key You will see a blinking character in the SrcA value field Confirm that...

Страница 16: ...A value field 2 Press the UP range arrow once so that the 10s digit is displayed It is displayed as 01 0000mA 3 Use the cursor key to highlight the 1s digit 4 Press the navigation wheel to enter EDIT...

Страница 17: ...ements begin Step 5 Observe measurements Observe the measured voltage on the main area of front panel display The readings should be very close to the 100 V source value Step 6 Turn output off When yo...

Страница 18: ......

Страница 19: ...priate to the final shipping destination of the instrument This line frequency is used for aperture NPLC calculations for the integrating analog to digital converter ADC You can manually configure the...

Страница 20: ...or to a safe high voltage SHV connector What should I do if I get an 802 interlock error You will receive error code 802 Output blocked by interlock if You disable the interlock when the Model 2657A o...

Страница 21: ...the Product Information CD ROM that came with your instrument Model 2657A Reference Manual Refer to the Model 2657A Reference Manual for detailed information about all features of the instrument This...

Страница 22: ......

Отзывы: