Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-200
2657A-901-01 Rev. B/December 2012
Usage
filterCount
= smu
X
.measure.filter.count
smu
X
.measure.filter.count =
filterCount
filterCount
The number of readings required for each filtered measurement (1 to 100)
X
Source-measure unit (SMU) channel (for example,
smua.measure.filter.count
applies to SMU channel A)
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
Sets the filter count for SMU channel A to 10.
Also see
(on page 7-200)
(on page 7-201)
(on page 7-211)
smuX.measure.filter.enable
This command enables or disables filtered measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
0 (smuX.FILTER_OFF)
Usage
filterState
= smu
X
.measure.filter.enable
smu
X
.measure.filter.enable =
filterState
filterState
The filter status; set to one of the following values:
0
or
smu
X
.FILTER_OFF
: Disables the filter
1
or
smu
X
.FILTER_ON
: Enables the filter
X
SMU channel (for example,
smua.measure.filter.enable
applies to SMU
channel A)
Details
This command enables or disables the filter.
Example
smua.measure.filter.enable = 1
Enables the filter for SMU channel A. Alternatively, the
value 1 may be replaced with
smua.FILTER_ON
.