Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-153
Details
If any parameters are omitted or
nil
, the operator is prompted to enter them using the front panel. Data for
pulsed current measurements, voltage levels, and timestamps are stored in
smu
X
.nvbuffer1
.
To perform the specified number of pulse V, measure I cycles, this function:
1. Sets the
smu
to output
bias
volts and dwell for
toff
seconds
2. Sets
the
smu
to output
level
volts and dwell for
ton
seconds
3. Performs voltage measurement with source at
level
volts
4. Sets
the
smu
to output
bias
volts for
toff
seconds
5. Repeats steps 2 through 4 for the remaining
points
pulse-measure cycles
Figure 116: PulseVMeasureI()
Example
smua.measure.nplc = 0.001
PulseVMeasureI(smua, -50, 125, 10e-3, 20e-3,
10)
SMU A outputs
−
50 V and dwells for 10 ms,
outputs 125 V and dwells for 20 ms. The
current measurement occurs during the
dwell period. After the measurement, the
output returns to
−
50 V and dwells for
10 ms. This pulse-measure process repeats
9 more times.
Also see
QueryPulseConfig()
This KIPulse factory script function allows you to inspect the settings of the preconfigured pulse train assigned to
tag
.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No
Usage
tbl
= QueryPulseConfig(
tag
)
tag
Numeric identifier to be assigned to the defined pulse train
tbl
Returned table
level
bias
ton
bias
toff
toff
Current measurement