Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-355
Also see
(on page 7-351)
(on page 7-355)
(on page 7-356)
(on page 7-358)
tsplink.trigger[N].pulsewidth
This attribute the length of time that the trigger line is asserted for output triggers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
TSP-Link trigger N reset
Not saved
10-e6
Usage
width
= tsplink.trigger[
N
].pulsewidth
tsplink.trigger[
N
].pulsewidth =
width
width
The pulse width (in seconds)
N
The trigger line (1 to 3)
Details
Setting the pulse width to 0 (seconds) asserts the trigger indefinitely.
Example
tsplink.trigger[3].pulsewidth = 20e-6
Sets pulse width for trigger line 3 to 20
μ
s.
Also see
(on page 7-355)
tsplink.trigger[N].release()