Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-284
2651A-901-01 Rev. A / March 2011
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the bit to set.
For example, to set bit B10, set
operationRegister
to 1024.
Example
status.operation.instrument.trigger_blender.enable = 1024
Sets the TRGOVR bit of the
operation status trigger
blender summary enable
using a decimal value.
Also see
status.operation.instrument.trigger_blender.trigger_overrun.*
(on page 7-284)
status.operation.instrument.trigger_blender.trigger_overrun.*
This attribute contains the operation status trigger blender overrun register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
30 (All bits set)
Usage
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.condition
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.enable
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.event
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.ntr
operationRegister
=
status.operation.instrument.trigger_blender.trigger_overrun.ptr
status.operation.instrument.trigger_blender.trigger_overrun.enable =
operationRegister
status.operation.instrument.trigger_blender.trigger_overrun.ntr =
operationRegister
status.operation.instrument.trigger_blender.trigger_overrun.ptr =
operationRegister
operationRegister
The operation status trigger blender overrun register's status. A zero (0) indicates no
bits set (also send 0 to clear all bits); other values indicate various bit settings