Bit
Value
Description
B0-B9
Not used
Not applicable
B10
status.operation.instrument.trigger_timer.TRIGGER_OVERRUN
status.operation.instrument.trigger_timer.TRGOVR
Set bit indicates
one or more
enabled bits for the
operation status
trigger timer
overrun register is
set.
Bit B10 decimal
value: 1,024
Binary value:
0100 0000 0000
B11-B15
Not used
Not applicable
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. For example, to set bit B10, set
operationRegister
to 1024.
Example
status.operation.instrument.trigger_timer.enable = 1024
Uses a decimal value to set
the TRGOVR bit of the
operation status trigger timer
summary enable register.
Also see
(on page E-7)
status.operation.instrument.trigger_timer.trigger_overrun.*
(on page 7-289)
status.operation.instrument.trigger_timer.trigger_overrun.*
This attribute contains the operation status trigger timer overrun register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
510 (All bits set)