2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-79
display.smuX.measure.func
This attribute specifies the type of measurement that is being displayed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Saved setup
1 (display.MEASURE_DCVOLTS)
Usage
func
= display.smu
X
.measure.func
display.smu
X
.measure.func =
func
func
The type of measurement:
0
or
display.MEASURE_DCAMPS
: Current measurement function
1
or
display.MEASURE_DCVOLTS
: Voltage measurement function
2
or
display.MEASURE_OHMS
: Resistance measurement function
3
or
display.MEASURE_WATTS
: Power measurement function
X
Source-measure unit (SMU) channel (for example,
display.smua.measure.func
applies to SMU channel A)
Details
Selects the measurement function that is displayed on the virtual front panel: Amps, volts, ohms, or
watts.
SMU A and SMU B can be set for different measurement functions.
Example
display.smua.measure.func = display.MEASURE_DCAMPS
Selects the current measure function
for SMU A.
Also see
display.trigger.clear()
This function clears the virtual front-panel trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last
display.trigger.wait()
call. This function clears the trigger event detector and discards the
previous history of TRIG key presses.
This attribute also clears the
display.trigger.overrun
attribute.