Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-78
2606B-901-01 Rev. B / May 2018
Usage
digits
= display.smu
X
.digits
display.smu
X
.digits =
digit
s
digits
Set digits to one of the following values:
Select 4-1/2 digit resolution (
4
or
display.DIGITS_4_5)
Select 5-1/2 digit resolution (
5
or
display.DIGITS_5_5)
Select 6-1/2 digit resolution (
6
or
display.DIGITS_6_5)
X
Source-measure unit (SMU) channel (for example,
display.smua.digits
applies to SMU channel A)
Details
SMU A and SMU B can be set for different measurement display resolutions.
Example
display.smua.digits = display.DIGITS_5_5
Select 5-1/2 digit resolution for SMU A.
Also see
(on page 3-67)
display.smuX.limit.func
If you are using a display mode that shows a single channel, this attribute specifies the type of limit value setting
displayed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Saved setup
0 (display.LIMIT_IV)
Usage
func
= display.smu
X
.limit.func
display.smu
X
.limit.func =
func
func
One of the following values:
0
or
display.LIMIT_IV
: Displays the primary limit setting
1
or
display.LIMIT_P
: Displays the power limit setting
X
Source-measure unit (SMU) channel (for example,
display.smua.limit.func
applies to SMU channel A)
Details
Selects the displayed limit function: primary (IV) or power (P).
SMU A and SMU B can be set for different display functions.
Example
display.smua.limit.func = display.LIMIT_P
Specifies the power limit value is
displayed for SMU Channel A.
Also see
(on page 7-79)
(on page 2-34)