22-10
Return to
2600AS-901-01 Rev. B / September 2008
Section 22: Performance Verification
Series 2600A System SourceMeter® Instruments Reference Manual
•
Select the correct source range.
•
Set the Series 2600A output current to the correct value.
•
Verify that the multimeter reading is within the limits given in the table.
6.
Repeat the procedure for negative output currents with the same magnitudes as those listed
in
.
7.
Turn the output off, and change connections as shown in
(use 4-wire
connections to the 0.5
Ω
resistor as shown).
8.
Select the DMM DC volts function.
9.
Repeat steps 4 through 7 for the 3A range (Model 2601A/2602A) or 1.5A range (Model
2611A/2612A/2635A/2636A). Calculate the current from the DMM voltage reading and
characterized 0.5
Ω
resistance value:
I=V/R.
10.
For the Model 2602A/2612A/2636A, repeat the above procedure for the other channel.
Table 22-6
Model 2601A/2602A output current accuracy limits
Model
2601A/2602A
source range
Model 2601A/2602A
output
current setting
Output current limits
(1 year, 18°C–28°C)
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
3A
90.000nA
0.90000µA
9.0000µA
90.000µA
0.90000mA
9.0000mA
90.000mA
0.90000A
2.40000A
89.846 to 90.154nA
0.89913 to 0.90087µA
8.9953 to 9.0047µA
89.943 to 90.057µA
0.89953 to 0.90047mA
8.9943 to 9.0057mA
89.953 to 90.047mA
0.89865 to 0.90135A
2.39706 to 2.40294A
Table 22-7
Model 2611A/2612A output current accuracy limits
Model
2611A/2612A
source range
Model 2611A/2612A
output
current setting
Output current limits
(1 year, 18°C–28°C)
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
90.000nA
0.90000µA
9.0000µA
90.000µA
0.90000mA
9.0000mA
90.000mA
0.90000A
1.35000A
89.846 to 90.154nA
0.89893 to 0.90107µA
8.9923 to 9.0077µA
89.913 to 90.087µA
0.89943 to 0.90057mA
8.9913 to 9.0087mA
89.943 to 90.057mA
0.89775 to 0.90225A
1.34519 to 1.35481A