5-6
Source-Measure Concepts
Model 2520 User’s Manual
Remote pulse parameters
Fixed mode
shows how various remote commands control various fixed pulse waveform
aspects. These parameters are programmed as follows:
•
SOUR1:CURR
— sets the pulse amplitude.
•
SOUR1:PULS :DEL
— sets the pulse delay (time between pulses).
•
SOUR1:PULS:WIDT
— sets the pulse width.
Figure 5-5
Remote fixed mode pulse parameters
Staircase sweep mode
shows how various front panel programming parameters control various stair-
case sweep pulse waveform aspects. These parameters are programmed by sending the
appropriate command:
•
Pulse amplitudes
— determined by parameters programmed with
SOUR1:CURR:STAR, SOUR1:CURR:STOP, and :SOUR1:CURR:STEP (or
SOUR1:SWE:POINT for LOG sweep) commands.
•
SOUR1:PULS:WIDT
— sets the pulse width.
•
SOUR1:PULS:DEL
— sets the pulse delay (time between pulses).
•
SOUR1:CURR:LOW
— This command sets the pulse low level. The pulse ampli-
tude returns to this level during the pulse delay unless the pulse amplitude is below
the low level, in which case the level is set to 0A.
SOUR1:CURR
SOUR1:PULS:WIDT
SOUR1:PULS:DEL