3-14
Basic Operation
Model 2520 User’s Manual
Programming example
summarizes the command sequence for basic source and measure configuration.
NOTE
See
for complete test procedures.
These commands set up the Model 2520 as follows:
•
Laser diode voltage measurement range and polarity: 10V, positive.
•
Photodiode measurement polarity: negative (both detectors).
•
Laser diode source: 0.5A range, +0.5A amplitude, 5V voltage limit, positive
polarity.
•
Laser diode source mode: pulse, delay = 100
µ
sec, width = 10
µ
sec, low I = 10mA.
•
Detector #1 source: +20V amplitude.
•
Detector #2 source: +10V amplitude.
Table 3-5
Basic source and measure configuration example
Commands
Comments
:SENS1:VOLT:RANG 10
:SENS1:VOLT:POL POS
:SENS2:CURR:POL NEG
:SENS3:CURR:POL NEG
:SOUR1:CURR:RANG 0.5
:SOUR1:CURR 0.5
:SOUR1:VOLT:PROT 5
:SOUR1:CURR:POL POS
:SOUR1:FUNC PULS
:SOUR1:PULS:DEL 100e-6
:SOUR1:PULS:WIDT 10e-6
:SOUR1:CURR:LOW 10e-3
:SOUR2:VOLT 20
:SOUR3:VOLT 10
Laser diode measure range = 10V.
Laser diode positive measure polarity.
Detector #1 negative measure polarity.
Detector #2 negative measure polarity.
0.5A laser diode source range.
Laser diode source output = 0.5A.
5V laser diode source voltage limit.
Laser diode source positive polarity.
Laser diode source pulse mode.
100
µ
sec pulse delay.
10
µ
sec pulse width.
10mA low current pulse.
Detector #1 bias source level = 20V.
Detector #2 bias source level = 10V.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...