Model 2520 User’s Manual
Common Commands
12-5
*SAV, *RCL programming example
summarizes the basic command sequence for saving and recalling a setup. The
present setup is stored in memory location 2, GPIB defaults are restored, and the memory
location 2 setup is recalled.
*RST — reset
Return Model 2520 to GPIB defaults
When the *RST command is sent, the Model 2520 performs the following operations:
•
Returns the Model 2520 to the GPIB default conditions. Refer to “Default parame-
ters” column of SCPI tables in
•
Cancels all pending commands.
•
Cancels response to any previously received *OPC and *OPC? commands.
*TRG — trigger
Send bus trigger to Model 2520
Use the *TRG command to issue a GPIB trigger to the Model 2520. It has the same effect
as a group execute trigger (GET).
Use the *TRG command as an event to control operation. The Model 2520 reacts to this
trigger if BUS is the programmed arm control source using the :TRIG:SOUR BUS com-
mand.
NOTE
Details on triggering are covered in
*TRG programming example
configures the Model 2520 to be controlled by bus
triggers. The last command, which sends a bus trigger, triggers one measurement. Each
subsequent bus trigger will also trigger a single measurement.
NOTE
With :TRIG:SOURce BUS selected, do not send any commands (except *TRG,
GET, DCL, SDC, IFC, and ABORt) while performing source-measure opera-
tions. Otherwise, erratic operation will occur.
Table 12-3
*SAV, *RCL programming example
Command
Description
*SAV 2
*RST
*RCL 2
Save present setup in memory location 2.
Restore GPIB defaults.
Recall location 2 setup.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...