9-4
Digital I/O Port, Interlocks, and Pulse Sync Output
Model 2520 User’s Manual
Source operation
shows the basic output configuration for source operation. In this case, the
external relay coil is connected between the digital output line (pins 1 to 4) and ground
(pin 9). With this configuration, the digital output line must be set HI to energize the relay,
and the maximum source current is 2mA.
Figure 9-3
Source operation
Controlling digital output lines
The Digital I/O lines can be used to control external relays or indicator lights. You can
control these lines either from the front panel or via remote as covered below.
Front panel digital output control
Set digital output line logic levels from the front panel as follows:
1.
Press the DIG OUT key.
2.
Using the EDIT keys, set the digital output parameter to the desired decimal value
(
). For example, to set the output lines to L, H, H, H, set the digital output
parameter value to 7.
3.
Press EXIT to return to normal display.
Model 2520
External
Relay
To Other
Circuits
+5V
Maximum Source Current: 2mA
Pin 7
Pins 1-4
Pin 9
Digital I/O
Port