7-6
Sweep Operation
Model 2520 User’s Manual
Performing sweeps
Procedures for the various sweep types are covered below.
NOTE
The following procedure assumes that the Model 2520 is already connected to
the DUT as explained in
Performing a linear staircase sweep
Step 1: Configure source and measure.
Configure the Model 2520 source and measure functions as follows:
1.
Select the current source range by pressing LASER I
L
then EDIT, then use the
RANGE keys.
2.
Press LASER V
L
, then choose the desired measurement range.
3.
Use the DETECTOR 1 V
B
and DETECTOR 2 V
B
keys, and EDIT keys to set the
photodiode bias voltages to the desired values.
4.
Set the current measurement ranges by pressing the DETECTOR 1 I
PD
, DETEC-
TOR 2 I
PD
, and RANGE keys.
See
, “
,” for more information.
Step 2: Configure sweep.
Configure the sweep as follows:
1.
Press CONFIG then SWEEP.
2.
Select TYPE, then press ENTER.
3.
Select STAIR, then press ENTER.
4.
At the prompts, enter the desired START, STOP, and STEP values.
5.
From the CONFIGURE SWEEPS menu, select SWEEP COUNT, press ENTER,
then choose FINITE or INFINITE as desired.
6.
Press EXIT to return to normal display.
Step 3: Turn outputs on.
Press the ON/OFF OUTPUT key to turn the outputs on (blue OUTPUT indicator turns
on). The Model 2520 will output the photodiode bias voltages.
Step 4: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the output off
by pressing the ON/OFF OUTPUT key.
Step 5: Read buffer.
Use the RECALL key to access the readings stored in the buffer.
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...