Model 2520 User’s Manual
Range, Filter, and Math
6-9
Remote math functions
Math function commands
summarizes commands to control the measurement math functions by remote.
See
“Calculate subsystems,” page 14-14
Table 6-9
Math function commands
Command
Description
:CALCulate[1]:DATA?
Request laser diode math reading.
:CALCulate[1]:FORMat <function>
Set laser diode math function (MXB[1],
CONDuctance[1], POWER[1], or
RESistance[1]).
:CALCulate[1]:KMATh:MBFactor <B>
Set laser diode MX + B offset (B) value.
:CALCulate[1]:KMATh:MMFactor <M>
Set laser diode MX + B slope (M) value.
:CALCulate[1]:KMATh:MUNits “<units>” Set 1-character MX + B units suffix.
:CALCulate[1]:STATe <state>
Enable/enable laser diode math (ON or OFF).
:CALCulate2:DATA?
Request detector #1 math reading.
:CALCulate2:KMATh:MBFactor <B>
Set detector #1 MX + B offset (B).
:CALCulate2:KMATh:MMFactor <M>
Detector #1 MX + B slope (M).
:CALCulate2:KMATh:MUNits “<units>”
Set 1-character MX + B units suffix.
:CALCulate2:STATe <state>
Enable/enable detector #1 math (ON or OFF).
:CALCulate3:DATA?
Request detector #2 math reading.
:CALCulate3:KMATh:MBFactor <B>
Set detector #2 MX + B offset (B).
:CALCulate3:KMATh:MMFactor <M>
Detector #2 MX + B slope (M).
:CALCulate3:KMATh:MUNits “<units>”
Set 1-character MX + B units suffix.
:CALCulate3:STATe <state>
Enable/enable detector #2 math (ON or OFF).
:CALCulate4:DATA?
Request delta (detector #1 current - detector
#2 current).
:CALCulate4:STATe <state>
Enable/disable delta (ON or OFF).
:INIT
Trigger CALC readings (must use before
sending CALCn:DATA?).
Содержание 2520
Страница 250: ...A Specifications ...
Страница 254: ...B StatusandErrorMessages ...
Страница 262: ...C DataFlow ...
Страница 266: ...D IEEE 488BusOverview ...
Страница 281: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 285: ...F MeasurementConsiderations ...
Страница 308: ...G GPIB488 1Protocol ...
Страница 312: ...H ExamplePrograms ...
Страница 317: ...I ContinuousPulseMode ...