7. Press ENTER. The instrument will prompt for calibration with -20V:
Ipd 1 CAL
Press ENTER to Output -20.000 V
8. Press ENTER. The unit will display:
DMM RDG: -20.000 V
Use ,
䊳
,
,
, ENTER or EXIT.
9. Use the EDIT keys to adjust the Model 2520 display value to agree with the DMM
voltage reading.
10. Press ENTER. The instrument will prompt for 0V calibration:
Ipd 1 CAL
Press ENTER to 00.000 V
11. Press ENTER. The unit will display:
DMM RDG: +00.000 V
Use ,
䊳
,
,
, ENTER or EXIT.
12. Use the EDIT keys to adjust the Model 2520 display value to agree with the DMM
reading.
13. Press ENTER. The instrument will prompt for calibration with +20V:
Ipd 1 CAL
Press ENTER to 20.000 V
14. Press ENTER. The unit will display:
DMM RDG: +20.000 V
Use ,
䊳
,
,
, ENTER or EXIT.
15. Use the EDIT keys to adjust the Model 2520 display value to agree with the DMM
reading, then press ENTER.
16. Press EXIT to return to normal display.
17. Repeat steps 3 through 16 for the 20mA, 50mA, and 100mA ranges, using
as a guide. Be sure to set the Model 2520 to the correct range using the
RANGE and keys, and use the correct resistor for each range.
18. Repeat steps 3 through 17 for the DETECTOR 2 measurement channel. Make
calibrator connections to the testhead DETECTOR 2 jack (see
Figure 2-8
Detector 2 current measurement calibration connections
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN
SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
CURRENT
OUTPUT
VOLTAGE
SENSE
CAT I
DETECTOR 1
DETECTOR 2
BIAS
CURRENT
INPUT
ISOLATION FROM EARTH: 10V MAX.
HI
HI
LO
LO
!
!
Model 2520 Testhead
Model 2001 DMM
Input
HI
Input LO
Inner
Shield
Center
Conductor
DETECTOR 2
Resistor
(See Text)
Содержание 2520
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 30: ......
Страница 31: ...2 Calibration...
Страница 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 69: ...3 RoutineMaintenance...
Страница 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 73: ...4 Troubleshooting...
Страница 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 91: ...5 Disassembly...
Страница 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 99: ...6 ReplaceableParts...
Страница 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 117: ...A Specifications...
Страница 121: ...B CalibrationReference...
Страница 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 136: ......