1-16
Performance Verification
Model 2520 Pulsed Laser Diode Test System Service Manual
Voltage bias source accuracy
Follow the steps below to verify that Model 2520 detector voltage bias source accuracy is
within specified limits. This test involves setting the bias voltages to specific values and
measuring the voltages with a DMM.
1. With the power off, connect the digital multimeter to the Model 2520 testhead
DETECTOR 1 jack, as shown in
.
2. Turn on the Model 2520 and DMM, and allow them to warm up for a minimum of
one hour.
3. Restore BENCH defaults as covered previously in
“Restoring factory defaults” on
.
4. Select the DMM DC volts measuring function, and choose auto-range.
Figure 1-9
Connections for DETECTOR 1 voltage bias source verification tests
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN
SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
CURRENT
OUTPUT
VOLTAGE
SENSE
CAT I
DETECTOR 1
DETECTOR 2
BIAS
CURRENT
INPUT
ISOLATION FROM EARTH: 10V MAX.
HI
HI
LO
LO
!
!
Model 2520 Testhead
Triax Cable
Model 2001 DMM
Input HI
Input LO
Center
Conductor
DETECTOR 1
Inner Shield
5. Press the DETECTOR 1 V
B
key, press EDIT, then set the source voltage to
+20.000V using the EDIT and or numeric keys.
6. Press the ON/OFF OUTPUT key to turn on the outputs, then press TRIG.
7. Verify the DMM reading is within the following limits: 19.75 to 20.25V.
8. Press the ON/OFF OUTPUT key to turn off the outputs.
9. Repeats steps 5 through 8 for the DETECTOR 2 bias source (make connections to
the testhead DETECTOR 2 jack as shown in
2 V
B
key to set the output voltage).
Содержание 2520
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 30: ......
Страница 31: ...2 Calibration...
Страница 68: ...2 38 Calibration Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 69: ...3 RoutineMaintenance...
Страница 72: ...3 4 Routine Maintenance Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 73: ...4 Troubleshooting...
Страница 90: ...4 18 Troubleshooting Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 91: ...5 Disassembly...
Страница 98: ...5 8 Disassembly Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 99: ...6 ReplaceableParts...
Страница 116: ...6 18 Replaceable Parts Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 117: ...A Specifications...
Страница 121: ...B CalibrationReference...
Страница 132: ...B 12 Calibration Reference Model 2520 Pulsed Laser Diode Test System Service Manual...
Страница 136: ......