1. Press the SET key to select the Output Settings Mode. A blinking cursor

will appear in the voltage field of the display.

2. Use the 

buttons to key in the desired output voltage

value, then press SET. The blinking cursor will move to the current field

of the display.

3. Use the 

buttons to key in the desired current limit, then

press SET to exit the Output Settings Mode. The display will return to

the actual V and I display mode.

GPIB commands for setting the output voltage and current limit:

:source:voltage xx.xxx

; set output voltage,

xx.xxx is between 0 and 20

:source:current:limit x.xxxx ; set output current limit,

x.xxxx is between 0 and 5

and

:sense:current:range minimum ; set to 5 mA range

or

:sense:current:range maximum ; set to 5 A range

2

Turning supply output on/off

The output of the Model 2304 is turned on and off using the front panel

OPERATE button. The word “ON” appears on the display when the output is turned

on. When the output is turned off, “OFF” is displayed.

GPIB commands for turning the supply output on/off:

:output on

; turn the supply output on

:output off

; turn the supply output off

M O D E L   2 3 0 4

4

Содержание 2304a

Страница 1: ...2304 MODEL 2304 QuickStart U S E R S G U I D E HIGHSPEED PRECISION READBACK P O W E R S U P P L Y...

Страница 2: ...acle Inspect the connecting cables test leads and jumpers for possible wear cracks or breaks before each use For maximum safety do not touch the product test cables or any other instruments while powe...

Страница 3: ...epend on the programming language used in the test environment Figure 1 Model 2304 front panel The Model 2304 has three operating display modes ACTUAL V AND I This mode reads back the actual output vo...

Страница 4: ...output voltage and current limit source voltage xx xxx set output voltage xx xxx is between 0 and 20 source current limit x xxxx set output current limit x xxxx is between 0 and 5 and sense current r...

Страница 5: ...of readings to average per measurement up to 10 optional command Figure 2 Rear panel of the Model 2304 Note the location of the digital voltmeter inputs 5 H I G H S P E E D P R E C I S I O N R E A D B...

Страница 6: ...f using the front panel controls press the DISPLAY button and select DVM INPUT using the and buttons GPIB commands for selecting the DVM Input Mode sense function dvmeter selects voltage measurement f...

Страница 7: ...a pulse The high measurement is triggered on the rising edge of the pulse and an integra tion is performed for the time specified for the high measurement The falling edge of the pulse triggers the lo...

Страница 8: ...t manually from 33 s to 833ms Make sure an integration period does not extend into the wrong portion of the pulse or into the next pulse For example if the pulse is high for 600 s the high integration...

Страница 9: ...OVERFLOW message will be displayed Perform the following procedure to find an appropriate trigger level Determining correct trigger level 1 Turn the output on If the trigger level is too low or too hi...

Страница 10: ...nhanced output for pulsed current measurements output on turn on supply output and sense pcurrent time high xxxxxx specify integration time in sec for high pulse measurements sense pcurrent time low x...

Страница 11: ...s pulsed current measurement function sense pcurrent synchronize tlevel x xxxx specify trigger level up to 5 amps sense pcurrent synchronize state off turn the trigger synchronization off sense pcurre...

Страница 12: ...0 248 6168 http www keithley com Copyright 1998 Keithley Instruments Inc Printed in the U S A Part No 2304 903 01 Rev A 1 98 F A X 440 248 6168 E MAIL product_info ke i t h l e y c o m S P E C D I R E...

Отзывы: