Specifications
A-3
PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL:
2303:
5mA to 5A, in 5mA steps.
2303-PJ: 5A Range:
0mA to 5A, in 5mA steps.
500mA Range:
0mA to 500mA, in 0.5mA steps.
TRIGGER DELAY:
0 to 100ms, in 10µs steps.
INTERNAL TRIGGER DELAY:
25µs.
HIGH/LOW/AVERAGE MODE:
Measurement Aperture Settings:
33.3µs to 833ms, in 33.3µs steps.
Average Readings:
1 to 100.
BURST MODE:
Measurement Aperture:
33.3µs.
Conversion Rate:
3600/second, typical.
Number of Samples:
1 to 5000.
Transfer Samples Across IEEE Bus in Binary Mode:
4800 bytes/ second, typical.
LONG INTEGRATION MODE: Measurement Time
6
:
850ms (840ms) to 60 seconds in 16.7ms (20ms) steps.
GENERAL
ISOLATION (low-earth):
22V DC max.
PROGRAMMING:
IEEE-488.2 (SCPI).
USER-DEFINABLE POWER-UP STATES:
5.
REAR PANEL CONNECTOR:
8-position quick disconnect terminal block for output (4), sense (2), and DVM (2).
TEMPERATURE COEFFICIENT (outside 23°C ±5°C):
Derate accuracy specification by (0.1
×
specification)/°C.
OPERATING TEMPERATURE:
0° to 35°C (Full power).
0° to 50°C (Derate to 70%).
STORAGE TEMPERATURE:
–20° to 70°C.
HUMIDITY:
<80% @ 35°C non-condensing.
POWER CONSUMPTION:
150VA max.
REMOTE DISPLAY/KEYPAD OPTION:
Disables standard front panel.
DIMENSIONS:
89mm high
×
213mm wide
×
360mm deep (3
1
⁄
2
in
×
8
1
⁄
2
in
×
14
3
⁄
16
in).
SHIPPING WEIGHT:
5.4kg (12 lbs).
INPUT POWER:
100V–240V AC, 50 or 60Hz (auto detected at power-up).
WARRANTY:
Two years parts and labor on materials and workmanship.
EMC:
Conforms with European Union Directive Directive 89/336/EEC EN 55011, EN 50082-1, EN 61000-3-2 and 61000-3-3, FCC part 15 class B.
SAFETY:
Conforms with European Union Directive 73/23/EEC EN 61010-1.
AC LINE LEAKAGE CURRENT:
450µA @ 110VAC, typ.; 600µA @ 220VAC, typ.
RELAY CONTROL JACK:
1-channel, sink 150mA max., 15V max. 5V output, 100mA max., also available on jack. Accepts 0.173 in Bantam-type
plug (CS-1003-1).
ACCESSORIES SUPPLIED:
User manual, service manual, output connector mating terminal (part no. CS-846).
ACCESSORIES AVAILABLE:
Model 2304-DISP:
Remote Display/Keypad (4.6 in
×
2.7 in
×
1.5 in). Includes 2.7m (9 ft) cable and rack mount kit.
Optional Version Model 2303B:
2303 with blank front panel (only AC power indicator LED).
1
PLC = 1.00.
2
Following 15 minute warm-up, the change in output over 8 hours under ambient temperature, constant load, and line operating conditions.
3
Remote sense, at output terminals, 1000% load change; typical.
4
Remote sense, with 4.5m (15 ft) of 16 gauge wire and 1
Ω
resistance in each source lead to simulate typical test environment, up to 1.5A load change.
5
Minimum current in constant current mode is 6mA.
6
60Hz (50Hz).
7
PLC = Power Line Cycle. 1PLC = 16.7ms for 60Hz operation, 20ms for 50Hz operation.
8
Display off.
9
Speed includes measurement and binary data transfer out of GPIB.
Specifications subject to change without notice.
Содержание 2303
Страница 28: ...1 14 Getting Started...
Страница 44: ...2 16 Basic Power Supply Operation...
Страница 64: ...4 10 Long Integration Measurements...
Страница 70: ...5 6 Relay Control...
Страница 102: ...7 20 Status Structure...
Страница 103: ...8 CommonCommands...
Страница 108: ...8 6 Common Commands...
Страница 109: ...9 Signal Oriented Measurement Commands...
Страница 121: ...11 SCPITables...
Страница 129: ...A Specifications...
Страница 132: ...A 4 Specifications...
Страница 133: ...B ErrorandStatusMessages...
Страница 137: ...C EmulationCommands...
Страница 143: ...D Models2303and2303B GPIB488 1Protocol...
Страница 151: ......