![Keithley 2281S Series Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/keithley/2281s-series/2281s-series_quick-start-manual_4046928023.webp)
1.
Press the
MENU
key.
2.
Highlight the
Settings
icon under Measure using either the
navigation control or the softkeys. Press
ENTER
.
3.
On the MEASURE SETTINGS page, select the button next to
Range
. The Range dialog box is displayed.
4.
Select the range. You are returned to the MEASURE
SETTINGS screen.
See the "General operation" section of the
Reference Manual
for
additional details and remote interface examples.
Application examples
Battery test
The following example demonstrates how to conduct a battery test
and generate a battery model with 2281S-20-6.
In this application, we will use the Model 2281S to charge a lithium-
ion battery from empty to full, and measure battery capacity (amp-
hour), battery resistance, charging curve (voltage/current), and finally
generate a battery model based on the test results.
For achieving a full-charge cycle, we need to use 2281S to discharge
the battery to an empty state first, then charge the battery from
empty to full state and measure voltage, current, resistance, and amp-
hour. After the charging process is done, the Model 2281S generate a
battery model automatically based on the results.
Equipment required:
One 2281S Battery Simulator
Rechargeable batteries
Wires
Model number
Current measurement
range
Voltage
measurement
range
2281S-20-6
10 A
20 V
1 A
100 mA
10 mA