Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-145
Example
TRAC:TRIG:CURR 1
TRAC:TRIG:CURR?
The current threshold value is set to 1.
Also see
:TRACe[n]:TRIGger:CURRent:STATe
(on page 7-145)
:TRACe[n]:TRIGger:CURRent:DIRection
(on page 7-144)
:TRACe[n]:TRIGger:CURRent:STATe
This command determines whether the current level trigger function is enabled.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0 (OFF)
Usage
:TRACe[n]:TRIGger:CURRent:STATe <state>
:TRACe[n]:TRIGger:CURRent:STATe?
<state>
ON
or
1
: Enable the current level trigger function
OFF
or
0
: Disable the current level trigger function
Example
TRAC:TRIG:CURR:STAT 1
TRAC:TRIG:CURR:STAT?3
The current level trigger function is enabled.
Also see
:TRACe[n]:TRIGger:CURRent[:LEVel]
(on page 7-144)
:TRACe[n]:TRIGger:CURRent:DIRection
(on page 7-144)