Series 2280 Precision DC Power Supply Reference Manual
Section 7: SCPI command reference
077085503 / March 2019
7-9
Example
*STB?
Query the contents of the Status Byte Register.
Output:
145
This output indicates that the Status Byte Register
contains the binary value
10010001
.
Also see
(on page D-10)
(on page D-1)
*TRG
This command generates a trigger event from a remote command interface.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
*TRG
Details
Use the
*TRG
command to generate a trigger event. Using
*TRG
has the same effect as a group
execute trigger (GET).
Use the
*TRG
command as an event to control operation. The instrument reacts to this trigger if
BUS
is the programmed arm control source. The control source is programmed from ARM and TRIGger
subsystems.
Example
*RST
:ARM:SOUR BUS
:ARM:COUN 1
:TRIG:SOUR IMM
:TRIG:COUN 1
:TRIG:SAMP:COUNT 1
:OUTP ON
:INIT
*TRG
Restore RST defaults.
Select BUS control source for arm layer.
Set arm layer count to 1.
Set the source trigger layer to trigger immediately.
Set trigger layer count to 1.
Set measure count to 1.
Turn on the output.
Take the Series 2280 out of the idle state.
Trigger one measurement.
Also see
(on page 7-18)
(on page D-1)
(on page 7-148)