Section 7: SCPI command reference
Series 2280 Precision DC Power Supply Reference Manual
7-6
077085503 / March 2019
*LANG?
This command determines which command set is used by the instrument.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Nonvolatile memory
SCPI
Usage
*LANG?
Details
The only available remote command set is SCPI, which is an instrument-specific language built on
the SCPI standard.
Example
*LANG?
Verify the setting by sending the language query.
Output:
SCPI
The SCPI command set is in use.
Also see
None
*RCL
This command returns the instrument to the setup that was saved with the
*SAV
command.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
*RCL <n>
[n]
An integer from 1 to 5 that represents the saved setup
Details
Restores the state of the instrument from a copy of its settings that are stored in the setup memory.
The settings are saved using the
*SAV
command.
Example
*RCL 3
Restores the settings stored in memory location
3.
Also see
(on page 7-7)
(on page 2-82)