OPERATION
15
Set to CV (Constant voltage) Mode
When setting the power supply to CV mode, a current limit must
also be set to determine the crossover point. When the current
exceeds the crossover point, the mode switches to CC mode. CC
and CV mode have two selectable slew rates: High Speed Priority
and Slew Rate Priority. High Speed Priority will use the fastest slew
rate for the instrument while Slew Rate Priority will use a user-
configured slew rate.
Background
Before setting the power supply to CV mode,
ensure:
•
The output is off.
•
The load is connected.
Steps
1.
Press the Function key. The
Function key will light up.
2.
The display should show F-
01 on the top and the
configuration setting for
F-01 on the bottom.
3.
Rotate the Voltage knob to change
the F setting to F-03 (V-I Mode
Slew Rate Select).
4.
Use the Current knob to set the F-
03 setting.
Set F-03 to 0 (CV High Speed
Priority) or 2 (CV Slew Rate
Priority).
F-03
0 = CV High Speed Priority
2 = CV Slew Rate Priority