2260B-GPIB-USB GPIB to USB Adapter
4
March 2014
Figure 2. The Measurement & Automation Explorer initial splash screen
3. From
the
Configuration panel access;
My System>Devices and Interfaces>GPIB0
4. Press
the
Scan for Instruments button.
5. In
the
Connected Instruments panel, the 2260B should be detected as Instrument 0 with the address the
same as that configured on the 2260B.
6. Double click the Instrument 0 icon.
7. Click
on
Communicate with Instrument.
8. In
the
NI-488.2 Communicator window, ensure
*IND?
is written in the Send String: text box.
Click on the Query button to send the
*IDN?
query to the instrument.
9. The
String Received text box will display the query return:
Keithley Instruments Inc.,Model 2260B-XX-XX, XXXXXX,XX.XX.XXXXXXXX
(manufacturer, model, serial number, version)
10. The function check is complete.