
Calibration considerations
When performing the calibration procedures:
•
Make sure that the test equipment is properly warmed up and connected to the
appropriate Model 2182 jack.
•
Always allow the source signal to settle before calibrating each point.
•
Use only copper-to-copper connections to minimize thermal EMFs.
•
Make sure that all connections are clean and free of oxidation to avoid thermal EMFs
that could affect calibration accuracy. Clean oxidation from connector terminals with
DeoxIT cleaning solution.
•
Keep test connections and all equipment away from drafts to avoid thermal drift.
•
Do not connect test equipment to the Model 2182 through a scanner or other switching
equipment.
•
If an error occurs during calibration, the Model 2182 will generate an appropriate error
message. See
Appendix B
for more information.
WARNING
The maximum voltage between terminals to chassis ground is 350V peak.
Exceeding this value may cause a breakdown in insulation, creating a shock
hazard.
CAUTION
The maximum voltage between CHANNEL 1 HI and LO is 120V. The
maximum voltage between CHANNEL 2 HI and LO is 12V. Exceeding
these voltage values may result in instrument damage.
Calibration cycle
Perform calibration every two years to ensure the unit meets or exceeds its specifications.
Calibration menu
The calibration menu allows you to access the calibration items summarized in
. To
access the calibration menu, press SHIFT then CAL. Use the RANGE and left and right arrow
keys to scroll through selections. Use the ENTER key to choose menu items, and use the EXIT
key to cancel them.
Table 2-1
Calibrating menu
Calibration menu item
Description
CAL: DATES
Display calibration and calibration due dates.
CAL: COUNT
Display number of times unit was calibrated.
CAL: RUN
Run calibration procedure.
CAL: TEMP
Display internal calibration temperature.
Calibration
2-3
Содержание 2182
Страница 10: ......
Страница 14: ...1 Performance Verification Performance Verification...
Страница 25: ...1 12 Performance Verification...
Страница 26: ...2 Calibration Calibration...
Страница 43: ...2 18 Calibration...
Страница 44: ...3 Routine Maintenance RoutineMain tenance...
Страница 47: ...3 4 Routine Maintenance...
Страница 48: ...4 Troubleshooting Troubleshoot ing...
Страница 64: ...5 Disassembly Disassembly...
Страница 73: ...5 10 Disassembly...
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ...6 Replaceable Parts Replaceable Parts...
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ...A Specifications Specifica tions...
Страница 96: ...B CalibrationReference Calibration Reference...
Страница 109: ...B 14 Calibration Reference...
Страница 110: ...C CalibrationProgram Calibration Program...
Страница 115: ...C 6 Calibration Program...
Страница 119: ......