Voltage and Temperature Measurements
2-3
Measurement overview
The Model 2182 provides two input channels for DC voltage and temperature measurements.
lists the measurements that can be performed by the two channels.
NOTE
Measurement queries are used to trigger and/or return readings. Details are provided
in
,
.
Channel 1 is used as the fundamental measurement channel, while Channel 2 provides sense
measurements. Because of this operational relationship between the two channels, Channel 2
cannot be used as an independent, stand-alone measurement channel. Its inputs must be
referenced to Channel 1 LO.
NOTE
As a general rule, use Channel 1 whenever possible for low voltage (<1V)
measurements. If using Channel 2 for measurements below 1V and the impedance
between Channel 2 LO and Channel 1 LO is
≥
100k
Ω
, pumpout current could be high
enough to corrupt measurements. For details, see “
Pumpout current (low charge injection mode)
Voltage measurements
The Model 2182 has two voltage measurement functions: DCV1 and DCV2. DCV1 is
available for input Channel 1, and DCV2 is available for Channel 2.
DCV1 (Channel 1) has five measurement ranges (10mV, 100mV, 1V, 10V, and 100V) and can
measure voltage from 1nV to 120V. DCV2 (Channel 2) has three measurement ranges (100mV,
1V, and 10V) and can measure voltage from 10nV to 12V. Accuracy for each channel is listed
in the specifications (
Temperature measurements
The Model 2182 has two temperature measurement functions: TEMP1 and TEMP2. TEMP1
is available for input Channel 1 and TEMP2 is available for Channel 2.
Depending on which thermocouple type is used (J, K, T, E, R, S, B, or N), the Model 2182
can measure temperature from -200˚C to 1820˚C. The specifications (
) provide the
measurement ranges for the various thermocouple types.
Table 2-1
Measurement channels
Measurement
Input Channel(s) To Use
Voltage
Channel 1
Temperature
Channel 1
Voltage and Voltage
Channel 1 and Channel 2
Voltage and Temperature
Channel 1 and Channel 2
Содержание 2182
Страница 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Страница 18: ......
Страница 22: ......
Страница 23: ...1 Getting Started Getting Started...
Страница 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Страница 68: ...2 28 Voltage and Temperature Measurements...
Страница 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Страница 82: ...3 14 Range Digits Rate and Filter...
Страница 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Страница 91: ...5 RatioandDelta Ratioand Delta...
Страница 117: ...6 Buffer Buffer...
Страница 123: ...7 Triggering Triggering...
Страница 140: ...7 18 Triggering...
Страница 141: ...8 Limits Limits...
Страница 149: ...9 SteppingandScanning Steppingand Scanning...
Страница 168: ...9 20 Stepping and Scanning...
Страница 169: ...10 AnalogOutput Analog Output...
Страница 175: ...11 RemoteOperation Remote Operation...
Страница 205: ...12 CommonCommands Common Commands...
Страница 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Страница 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Страница 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Страница 260: ...15 22 Additional SCPI Commands...
Страница 261: ...A Specifications Specifications...
Страница 263: ...B Statusand ErrorMessages StatusandError Messages...
Страница 268: ...B 6 Status and Error Messages...
Страница 269: ...C Measurement Considerations Measurement Consider ations...
Страница 278: ...C 10 Measurement Considerations...
Страница 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Страница 284: ...D 6 Model 182 Emulation Commands...
Страница 285: ...E Example Programs ExamplePro grams...
Страница 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Страница 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Страница 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Страница 311: ...H Measurement Queries Measurement Queries...
Страница 316: ...H 6 Measurement Queries...
Страница 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...