2.
Turn on the power, and allow a 2-1/2 hour warm-up period.
3.
Restore factory defaults, and do a full auto-calibration, as previously described.
4.
Select the Model 2182 DCV1 measurement function, and select the 100mV range.
5.
Set the calibrator output to 0.00000mV, and allow the reading to settle.
6.
Enable REL, and leave it enabled for the remainder of the tests.
7.
Verify reading accuracy for each of the voltages listed in
. For each test point:
•
Select the correct Model 2182 measurement range.
•
Set the DC calibrator output voltage to the indicated value.
•
Verify that the Model 2182 voltage reading is within the limits given in the table.
8.
Repeat step 7 for negative source voltages with the same magnitudes as those listed in
.
9.
Repeat steps 4 through 8 for the 100mV-10V ranges of the CHANNEL 2 input. Be sure
to connect the calibrator to the appropriate input jack terminals by changing input cable
wire connections, and select the DCV2 measurement function.
10mV range accuracy
1.
With the power off, connect the DC calibrator and low-thermal divider to the
Model 2182 CHANNEL 1 terminals of the input jack, as shown in
Figure 1-2
Connections for 10mV range accuracy
2.
Turn on the power, and allow a 2-1/2 hour warm-up period.
3.
Select the Model 2182 DCV1 measurement function, and choose the 10mV range.
4.
Select the 100:1 divider ratio.
5.
Set the calibrator output to 0.00000V, and allow the reading to settle.
6.
Enable REL.
!
CHANNEL 1
CHANNEL 2
Model 2182
OUTPUT
SENSE
HI
LO
INPUT
2107 Low
Thermal Cable
LO
HI
Wrap Connections
in Foam Insulation
Model 262
Low Thermal
Divider
Copper #18 AWG
insulated wire (Solid
or Stranded)
SENSE
LO
OUTPUT
LO
OUTPUT
HI
SENSE HI
DC Voltage Calibrator
1506 Low
Thermal Cable
Performance Verification
1-7
Содержание 2182
Страница 10: ......
Страница 14: ...1 Performance Verification Performance Verification...
Страница 25: ...1 12 Performance Verification...
Страница 26: ...2 Calibration Calibration...
Страница 43: ...2 18 Calibration...
Страница 44: ...3 Routine Maintenance RoutineMain tenance...
Страница 47: ...3 4 Routine Maintenance...
Страница 48: ...4 Troubleshooting Troubleshoot ing...
Страница 64: ...5 Disassembly Disassembly...
Страница 73: ...5 10 Disassembly...
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ...6 Replaceable Parts Replaceable Parts...
Страница 89: ......
Страница 90: ......
Страница 91: ......
Страница 92: ...A Specifications Specifica tions...
Страница 96: ...B CalibrationReference Calibration Reference...
Страница 109: ...B 14 Calibration Reference...
Страница 110: ...C CalibrationProgram Calibration Program...
Страница 115: ...C 6 Calibration Program...
Страница 119: ......