:CONFigure:<function>
[<range>|MIN|MAX|DEF,<resolution>|MIN|MAX|DEF]
This command configures the instrument to a specific setup for a specified measurement function. This command
does not initiate the measurement. The :READ? command can be used to trigger a measurement. You can also
use the INIT;:FETC? command.
Parameters
All <range> and <resolution> parameters also include MIN, MAX, and DEF. N/A indicates that the
<range> and/or <resolution> parameters are not applicable for the <function>.
<function>
<range>
<resolution> (see note 1)
Description (see
note 2)
CURRent[:DC]
-10 to 10
3.00e-07 to 3.00e-04
Amperes
CURRent:AC
0 to 10
1.00e-06 to 1.00e-04
Amperes
VOLTage[:DC]
-1000 to 1000
3.00e-07 to 3.00e-04
Voltage
VOLTage:AC
0 to 750
1.00e-06 to 1.00e-04
Voltage
CAPacitance
0 to 10e-3
N/A
Farads
VOLTage[:DC]:RATio
-1000 to 1000
3.00e-07 to 3.00e-04
Voltage
RESistance
0 to 100e6
3.00e-07 to 3.00e-04
Ohms
FRESistance
0 to 100e6
3.00e-07 to 3.00e-04
Ohms
FREQuency[:VOLTage]
0 to 750
N/A
Voltage
FREQuency:CURRent
0 to 10
N/A
Amperes
PERiod[:VOLTage]
0 to 750
N/A
Voltage
PERiod:CURRent
0 to 10
N/A
Amperes
CONTinuity
N/A
N/A
N/A
DIODe
N/A
N/A
N/A
TCOuple
N/A
N/A
N/A
TEMPerature
N/A
N/A
N/A
1. Do not use the number of digits to specify a resolution; instead, use the same units as the
measurement function. For example, specify the resolution in volts and the frequency in hertz. You
must multiply the resolutions in the table by the needed range before using them as resolution
arguments. Refer to
Bandwidth and resolution for AC measurements
(on page 3-44) for more
information about AC measurements. Refer to
Integration time and resolution for DC and resistance
(on page 3-41) for information about DC measurements.
2. Information listed in the description column is for reference only. This column lists the
measurement unit for the <function>. Do not include units in commands.