A-5
AC VOLTS (cont’d)
NORMAL MODE RMS
1
90 Days,
±
2
°
C from last AC self-cal for 1% to 100% of range
3
±
(% of r % of range)
RANGE
20–50Hz
50–100Hz
0.1–2kHz
2–10kHz
10–30kHz
30–50kHz
50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV
0.25+0.015
0.07+0.015
0.03+0.015
0.03+0.015
0.035+0.015
0.05+0.015
0.17+0.015
0.5+0.025
2+0.1
5+0.2
2
V
0.25+0.015
0.07+0.015
0.03+0.015
0.03+0.015
0.035+0.015
0.05+0.015
0.17+0.015
0.5+0.025
2+0.1
5+0.2
20
V
0.25+0.015
0.07+0.015
0.04+0.015
0.06+0.015
0.08 +0.015
0.1 +0.015
0.17+0.015
0.5+0.025
4+0.2
7+0.2
5
200
V
4
0.25+0.015
0.07+0.015
0.04+0.015
0.06+0.015
0.08 +0.015
0.1 +0.015
0.17+0.015
0.5+0.025
5
4+0.2
5
750
V
4
0.25+0.015
0.1 +0.015
0.08+0.015
0.09+0.015
0.12 +0.015
0.15+0.015
5
0.5 +0.015
5
NORMAL MODE RMS
1
1 or 2 Years, T
CAL
±
5
°
C for 1% to 100% of range
3
±
(% of r % of range)
RANGE
20–50Hz
50–100Hz
0.1–2kHz
2–10kHz
10–30kHz
30–50kHz
50–100kHz 100–200kHz 0.2–1MHz 1–2MHz
200 mV
0.25+0.015
0.08+0.015
0.05+0.015
0.05 +0.015
0.05+0.015
0.06+0.015
0.17+0.015
0.5+0.025
2+0.1
5+0.2
2
V
0.25+0.015
0.08+0.015
0.05+0.015
0.05 +0.015
0.05+0.015
0.06+0.015
0.17+0.015
0.5+0.025
2+0.1
5+0.2
20
V
0.25+0.015
0.08+0.015
0.06+0.015
0.085+0.015
0.12+0.015
0.13+0.015
0.17+0.015
0.5+0.025
4+0.2
7+0.2
5
200
V
4
0.25+0.015
0.08+0.015
0.06+0.015
0.085+0.015
0.12+0.015
0.13+0.015
0.17+0.015
0.5+0.025
5
4+0.2
5
750
V
4
0.27+0.015
0.11+0.015
0.1 +0.015
0.13 +0.015
0.18+0.015
0.22+0.015
5
0.5 +0.015
5
dB ACCURACY RMS
±
dB, 90 Days, 1 or 2 Years, T
CAL
±
5
°
C, Reference=1V, Autoranging, Low Frequency Mode, AC+DC Coupling
INPUT
1–100Hz
0.1–30kHz
30–100kHz
100–200kHz
0.2–1MHz
1–2MHz
–54 to–40
dB
(2 mV to
10 mV)
0.230
0.225
0.236
0.355
–40 to–34
dB
(10 mV to
20 mV)
0.036
0.031
0.041
0.088
–34 to
6
dB
(20 mV to
2
V)
0.023
0.018
0.028
0.066
0.265
0.630
6 to 26
dB
(2
V to
20
V)
0.024
0.024
0.028
0.066
0.538
0.820
5
26 to 46
dB
(20
V to 200
V)
0.024
0.024
0.028
0.066
5
0.538
5
46 to 57.8 dB (200
V to 775
V)
0.018
0.021
0.049
5
ACV READING RATES
5,6
READINGS/SECOND WITH
MEASUREMENT
DEFAULT
READINGS/SECOND TO MEMORY
READINGS/SECOND TO IEEE-488
TIME STAMP TO IEEE-488
NPLC
APERTURE
BITS
DIGITS
Auto Zero Off
Auto Zero On
Auto Zero Off
Auto Zero On
Auto Zero Off Auto Zero On
10
167 ms (200 ms)
28
6
1
⁄
2
6
(5.1)
2
(1.7)
2
2
(1.6)
2
2
(1.5)
2
33.4 ms
(40 ms)
26
5
1
⁄
2
30
(24)
9
(7.9)
28
(23)
9
(7.6)
27
(22)
9
(7.5)
1
16.7 ms
(20 ms)
25
5
1
⁄
2
57
(48)
38
(35)
53
(45)
36
(33)
48
(41)
34
(30)
0.1
1.67 ms
(2 ms)
21
5
1
⁄
2
136 (136)
70
(70)
122 (122)
64
(64)
98
(98)
56
(56)
0.01
167
µ
s (167
µ
s)
16
4
1
⁄
2
140 (140)
71
(71)
127 (127)
66
(66)
99
(99)
58
(58)
0.01
8
167
µ
s (167
µ
s)
16
4
1
⁄
2
2000 (2000)
2000 (2000)
AC COUPLING:
For AC only coupling, add the following % of reading:
1–10Hz
10–20Hz
20–50Hz 50–100Hz 100–200Hz
Normal Mode
(rms, average)
—
—
0.41
0.07
0.015
Low Frequency Mode
(rms)
0.1
0.01
0
0
0
For low frequency mode below 200Hz, specifications apply for sine wave inputs only.
AC+DC COUPLING:
For DC>20% of AC rms voltage, apply the following
additional uncertainty, multiplied by the ratio (DC/AC rms). Applies to rms
and average measurements.
RANGE
% of Reading
% of Range
200mV, 20V
0.05
0.1
2V, 200V, 750V
0.07
0.01
ACV CREST FACTOR MEASUREMENT
11
CREST FACTOR
= Peak AC / rms AC.
CREST FACTOR RESOLUTION:
3 digits.
CREST FACTOR ACCURACY:
Peak AC uncer AC normal mode rms
uncertainty.
MEASUREMENT TIME:
100ms plus rms measurement time.
INPUT CHARACTERISTICS:
Same as ACV input.
CREST FACTOR FREQUENCY RANGE:
20Hz – 1MHz.
CREST FACTOR DISPLAY:
Access as multiple display on AC volts.
ACV PEAK VALUE MEASUREMENT
10
REPETITIVE PEAK ACCURACY,
±
(% of % of range), 90 Days, 1 Year or 2 Years, T
CAL
±
5
°
C
TEMPERATURE COEFFICIENT
20Hz–
1kHz–
10kHz–
30kHz–
50kHz–
100kHz–
300kHz–
500kHz–
750kHz–
±
(% of % of range)/
°
C
RANGE
1kHz
9
10kHz
30kHz
50kHz
100kHz
300kHz
500kHz
750kHz
1MHz
Outside T
CAL
±
5
°
C
200 mV
0.08+0.7
0.09+0.7
0 .1 +0.7
0.15+0.7
0.25+0.7
1.0+0.7
2.5+0.7
5.5+0.7
9+0.7
0.002 + 0.03
2
V
0.08+0.3
0.09+0.3
0 .1 +0.3
0.15+0.3
0.25+0.3
1.0+0.3
2.5+0.3
5.5+0.3
9+0.3
0.002 + 0.03
20
V
0.1 +0.7
0.11+0.7
0.14+0.7
0.19+0.7
0.25+0.7
1.0+0.7
2.5+0.7
5.5+0.7
9+0.7
0.004 + 0.03
200
V
4
0.1 +0.3
0.11+0.3
0.14+0.3
0.19+0.3
0.25+0.3
1.0+0.3
5
2.5+0.3
5
5.5+0.3
5
9+0.3
5
0.004 + 0.03
750
V
4
0.12+0.6
0.16+0.6
0.2 +0.6
0.25+0.6
5
0.5 +0.6
5
0.01 + 0.02
Valid % of Range
7
10
–
400%
10
–
400%
10
–
400%
10
–
350%
10
–
350%
10
–
250%
10
–
150%
10
–
100%
7.5
–
75%
PEAK MEASUREMENT WINDOW:
100ms per reading.
MAXIMUM INPUT:
±
1100V peak, 2
×
10
7
V•Hz (for inputs above 20V).
DEFAULT MEASUREMENT RESOLUTION:
4 digits.
NON-REPETITIVE PEAK:
10% of range per
µ
s typical slew rate for single spikes.
PEAK WIDTH:
Specifications apply for all peaks
≥
1
µ
s.
AVERAGE ACV MEASUREMENT
Normal mode rms specifications apply from 10% to 100% of range, for 20Hz–
1MHz. Add 0.025% of range for 50kHz–100kHz, 0.05% of range for 100kHz–
200kHz, and 0.5% of range for 200kHz–1MHz.
HIGH CREST FACTOR ADDITIONAL ERROR
±
(% of reading)
Applies to rms measurements.
CREST FACTOR:
1 – 2
2 – 3
3 – 4
4 – 5
ADDITIONAL ERROR:
0
0.1
0.2
0.4
Содержание 2001
Страница 1: ...Model 2001 Multimeter Repair Manual Contains Servicing Information...
Страница 129: ...Troubleshooting 2 116...
Страница 139: ...Replaceable Parts 4 4...
Страница 149: ...Replaceable Parts 4 14...
Страница 153: ...Replaceable Parts 4 18...
Страница 156: ...A Model 2001 Specifications A 1...