75
:VOLTage:DC
:AVERage
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe] <b>
[:STATe]?
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
Path to configure and con-
trol advanced filter.
Specify noise tolerance
level (0 to 100 percent).
Query noise tolerance
level.
Enable (1 or ON) or dis-
able (0 or OFF)
advanced filter.
Query state of advanced
filter.
Enable (1 or ON) or disable
(0 or OFF) filter.
Query state of digital filter.
Enable (1 or ON) or disable
(0 or OFF) auto filter.
Enable and then disable
auto filter.
Query auto filter.
:RESistance
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
:RANGe
Path to configure resistance:
Specify integration rate in sec-
onds (166.67e-6 to 200e-3).
Enable (1 or ON) or disable
(0 or OFF) auto aperture.
Enable and then disable
auto aperture.
Query auto aperture (0 or 1).
Query aperture (integration
rate).
Specify integration rate in
number of line cycles (0.01
to 10).
Enable (1 or ON) or disable
(0 or OFF) auto NPLC.
Enable and then disable
auto NPLC.
Query auto line cycle inte-
gration (0 or 1).
Query line cycle integration
rate.
Path to configure measure-
ment range:
Table 29. Sense command summary (cont.)
Command
Description