In this section:
Introduction .............................................................................. 3-1
Signal limitation ........................................................................ 3-1
Scanner card detection ............................................................ 3-2
Basic front-panel measurements .............................................. 3-2
Typical applications .................................................................. 3-7
Measurement considerations ................................................. 3-14
Introduction
This section contains detailed information on front panel and remote interface operation of the
2001-TCSCAN.
Before using the 2001-TCSCAN, you should be thoroughly familiar with the operation of the
DMM6500 multimeter. See the DMM6500 User's and Reference Manuals for the necessary
information.
Signal limitation
To prevent overheating or damage to the relays, never exceed the following maximum input signal
levels:
•
DC signals
110 V DC
1 A switched
30 VA (resistive load)
•
AC signals
125 V
RMS
or 175 V AC
PEAK
1 A switched
62.5 VA (resistive load)
Channels 5 and 10 have a maximum power of 0.25 VA due to factory-installed, current-limited
resistors.
Before using the 2001-TCSCAN, you should be thoroughly familiar with the operation of the
DMM6500 multimeter. See the DMM6500 User's and Reference Manuals for more information.
Section 3
Operation