![Keithley 2000-Scan Скачать руководство пользователя страница 22](http://html2.mh-extra.com/html/keithley/2000-scan/2000-scan_user-manual_3934049022.webp)
Section 2: Card connections and installation
Model 2000-SCAN Scanner Card User's Manual
2-14
2000-SCAN-900-01 Rev. A / April 2018
To connect the output leads for 4-pole connection:
1. Connect OUT A HI (red) to INPUT HI.
2. Connect OUT A LO (black) to INPUT LO.
3. Connect OUT B HI (red) to SENSE HI.
4. Connect OUT B LO (black) to SENSE LO.
See the following figure.
Figure 11: 4-pole output connections
Scanner card removal
Turn off power to all instrumentation, including the multimeter, and disconnect all line cords.
Make sure all power is removed and any stored energy in external circuitry is discharged.
To remove the scanner card from the multimeter:
1. Unlock the scanner card by pulling the ejector arms outward.
2. Carefully slide the scanner card out of the multimeter.
3. If you will operate the DMM6500 without the scanner card installed, install the cover plate over
the Accessory Card Slot.