3-10
Operation
4-wire resistance tests
More precise measurements over a wider range of system and DUT conditions can be
obtained using the 4-wire measurement scheme shown in
from the Model 2000-20 Multimeter are routed through the scanner to the resistor under test.
The extra set of sense leads minimizes the effects of voltage drops across the test leads, greatly
reducing measurement errors with lower DUT resistances. Note, however, that an extra two
poles of switching are required for each resistor tested. For this reason, only ten resistors can be
tested using this configuration.
Figure 3-2
4-wire resistance test connections
DMM 2000-20
Rear Panel
1
DUTs
(10)
R
A. Test Configuration
B. Simplified Equivalent Circuit
10
11
20
1
10
HI
LO
Sense LO
HI
Sense HI
LO
Sense LO
2000-20
2000-SCAN-20
DUT
2000-SCAN-20
Out A
Out B
Sense HI
Ω
Содержание 2000-SCAN-20
Страница 11: ......
Страница 12: ...1 General Information General Information ...
Страница 17: ...1 6 General Information ...
Страница 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Страница 35: ...2 18 Card Connections and Installation ...
Страница 36: ...3 Operation Operation ...
Страница 51: ...3 16 Operation ...
Страница 52: ...4 Service Information Service Information ...
Страница 63: ...4 12 Service Information ...
Страница 64: ...5 Replaceable Parts Replace ableParts ...
Страница 67: ...5 4 Replaceable Parts ...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Страница 79: ......
Страница 80: ......