Card Connections and Installation
2-13
Current connections
shows typical connections for measuring currents using the
Model 2000-SCAN-20 with current shunts installed. As noted previously, only channels 1 and
11 have current shunts installed. This configuration can be used to scan the following:
•
DC current
•
AC current
Figure 2-8
Current connections
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
HI
LO
CH 1
CH 2
CH 10
CH 11
CH 12
CH 20
OUT A
To DMM Input
2000-SCAN-20 Card
AC or DC
Current
HI
LO
HI
LO
OUT B
J1036
J1035
A
A
AC or DC
Current
NOTE: Current shunts
must be installed
for channels 1
and 11.
Содержание 2000-SCAN-20
Страница 11: ......
Страница 12: ...1 General Information General Information ...
Страница 17: ...1 6 General Information ...
Страница 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Страница 35: ...2 18 Card Connections and Installation ...
Страница 36: ...3 Operation Operation ...
Страница 51: ...3 16 Operation ...
Страница 52: ...4 Service Information Service Information ...
Страница 63: ...4 12 Service Information ...
Страница 64: ...5 Replaceable Parts Replace ableParts ...
Страница 67: ...5 4 Replaceable Parts ...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Страница 79: ......
Страница 80: ......