11
Press LOCAL to cancel remote (REM on) when the
instrument is used over the IEEE-488 bus. LOCAL
will be locked out when the bus LLO command is in
effect.
FACTORY DEFAULT CONDITIONS
Description
Default State
Range
30V, auto-range off
Reading relative
Off, value =0V
Analog Filter
Off
Digital Filter
On, medium response
Buffer
Off, size = 0000 (wrap around)
Analog Output
Relative off, value: 0V, gain: 1.0,
normal output, source value
= 0V
Primary Address
7
Integration Period
Line cycle (16.67msec @ 60Hz,
20msec @ 50Hz)
Display Resolution 6 1/2 digits
Trigger Mode
Multiple
Trigger Source
External (MANUAL key also
operative)
Trigger Interval
0.25sec
Trigger Delay
0sec
Calibration Lock
On (calibration locked)
NOTE: All listed operating states except buffer status, primary address,
and calibration lock status can be saved as user configuration using
MENU save user setup selection. Primary address can be saved using
MENU SET IEEE ADDRESS selection.