MFS 

 

 

I

NSTALLATION

 

2.1  Unpacking & Handling 

 

When unpacking or handling the oven always lift it by its case.  Do not apply lifting force to a 
door or any other protruding part. 

 

Remove any packing material from the oven chamber(s) before use. 

2.2 

Siting & Setting Up 

 

Place the oven in a well ventilated room, away from other sources of heat, and on a surface which 
is resistant to accidental spillage of hot materials.  Do not mount the oven on an inflammable 
surface. 

 

Ensure that there is free space around the oven.  Do not obstruct any of the vents in the case: they 
are needed to keep the controls and the case exterior cool. 

 

Ensure that the oven is placed in such a way that it can be quickly switched off or disconnected 
from the electrical supply - see below. 

 

Each chamber is fitted with an air guide (a horizontal plate dividing the top from the bottom), and 
a sample tray (in the lower part of the chamber). Ensure that these items are present. Do not 
remove the guide - it is required for correct gas flow within the unit. 

2.3 

Electrical Connections 

 

Connection by a qualified electrician is recommended.

 

 

The oven requires a single-phase A.C. supply with earth (ground).  The supply may be Live to 
Neutral non-reversible, Live to Neutral with reversible plug, or Live to Live. 

 

The voltage or range of voltages on which the oven may be operated is given on the oven rating 
label and on the cover of this manual (if supplied with the oven).  Check that the supply voltage is 
compatible with the voltage on the label, and that the current capacity is sufficient for the 
amperage on the label, before connection to the supply.  See also section 8.0. 

 

The oven is normally supplied fitted with a supply cable; it may be wired directly to the supply 
point, which must be within reach of the oven operator and must incorporate either an isolating 
switch which operates on both conductors, or a quickly removable plug. 

 

The supply MUST incorporate an earth (ground). 

 

Single phase: connect the supply as follows: 

 

 

supply type

           

 

Cable colour 

Terminal 
 label 

Live-
Neutral 

Reversible or 
Live-Live 

Brown 

    L 

to live 

to either power 
conductor 

Blue 

    N 

to neutral 

to the other power 
conductor 

Green/Yellow 

    PE 

to earth 

to earth 

 

2.4 

Gas Connections (MFS/1) 

 

The upper gas connection at the back of the oven is the nitrogen inlet, supplying gas to the left 
hand flowmeter whose valve may be adjusted to set the nitrogen flow rate in accordance with the 
requirements of the test standard. 

 

The lower gas connection is the combined outlet from the two exhaust paths from the chamber. 

 

The first of those paths goes from the chamber through the right hand flowmeter to the exhaust 
port. 

MF14 

Содержание Carbolite MFS/1

Страница 1: ... oven is connected to the electricity supply CONTENTS Section page 1 0 Symbols Warnings 2 2 0 Installation 3 3 0 Operation 5 4 0 Maintenance 6 5 0 Repairs Replacements 7 6 0 Fault Analysis 9 7 0 Circuit Diagrams 10 8 0 Fuses Power Settings 10 9 0 Specifications 12 A manual is supplied separately for the temperature controller Please read the controller manual before operating the oven 1 0 MF14 ...

Страница 2: ... Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements 1 2 Warning Symbols DANGER of electrical shock read any warning printed by this symbol DANGER hot surface Read any warning printed by this symbol WARNING all surfaces of a furnace may be hot DANGER read any warning printed by this symbol 2 0 2 MF14 ...

Страница 3: ...utral with reversible plug or Live to Live The voltage or range of voltages on which the oven may be operated is given on the oven rating label and on the cover of this manual if supplied with the oven Check that the supply voltage is compatible with the voltage on the label and that the current capacity is sufficient for the amperage on the label before connection to the supply See also section 8...

Страница 4: ...tlet for all four chambers on the right hand side of the oven By switching one of the four valves on the right hand side of the oven labelled ____ TEST to the TEST position the outlet from the corresponding chamber passes through the flowmeter labelled TEST This allows a comparison of inward and outward gas flows for that chamber to test for leaks Each chamber can be tested in turn To avoid moistu...

Страница 5: ...oller see the instructions in the separate manual supplied Overtemperature hydraulic thermostat MFS 4 if fitted Set the rotary dial to the desired protection temperature The oven starts to heat up according to the controller set point or program To switch the oven off set the Instrument switch to off If the oven is to be left off isolate it from the electrical supply 3 2 MFS 4 Chamber Temperatures...

Страница 6: ... Carbolite In all correspondence please quote the serial number model type and voltage given on the rating label of the furnace The serial number and model type are also given on the front of this booklet when supplied with a furnace To contact Thermal Engineering Services or Carbolite see the back page of this manual 4 4 Recommended Spares Kits Carbolite can supply individual spares or a kit of t...

Страница 7: ...6 We can provide further information on request Alternatively our service division can quote for any repairs to be carried out at your premises or ours 5 3 Temperature Controller Replacement 200 201 These controllers are fitted to the back of the control panel which can be separated from the base by removal of two screws Before handling the controller wear an anti static wrist strap or otherwise a...

Страница 8: ...the new thermocouple bending to shape as required to match the old Reconnect and refit the panels 5 7 Element Replacement MFS 1 See section 5 2 wearing a face mask is recommended Disconnect the oven from the electrical supply and remove the back panel Make a note of the wiring and piping connections see the section above which refers to the thermocouple colour coding Disconnect terminal blocks the...

Страница 9: ...thermocouple has broken or has a wiring fault Æ The controller shows a low temperature Æ The SSR could be failing to switch on due to internal failure faulty logic wiring form the controller or faulty controller Æ There are no lights glowing on the controller Æ The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault B Oven Overheats 1 Oven only heats up when...

Страница 10: ... e m e n t s F1 N L PE thermocouple Supply light Instrument Switch temperature controller PE L N F1 SSR e l e m e n t s F2 Temperature Sensor Hydraulic thermostat Warning light Hydraulic t stat o temp if fitted Reset button Fan motor temperature indicator Chamber Thermocouples Filter Filter 8 0 10 MF14 ...

Страница 11: ...32mm x 6mm type F1 quick blow F2 Instrument circuit fuses 20mm x 5mm glass 2 Amps type F1 omitted in MFS 1 model volts phases supply fuse rating MFS 1 220 240V 1 2A MFS 4 220 240V 2 10A 8 2 Power Settings For all models in this manual the power limit parameter OP Hi is set to 100 MF14 11 ...

Страница 12: ...l 41 41 541 5877 Fax 41 41 541 5878 Email General Information Sales info kesn ch Accounts Department accounts kesn ch USA Office 233 Rogue River Highway 425 Grants Pass Oregon 97527 5429 USA Tel 1 503 206 0941 Fax 1 541 610 1646 Email General Information Sales info keison com Accounts Department accounts keison com Latin America Office World Trade Centre Apartado 0832 00708 Ciudad de Panamá República ...