background image

SPECIFICATION 

 
Model: KANE3500-30 
 

Nominal Pressure Ranges 

 

mBar

 

kPa

 

PSI

 

 

2070.0

 

207.00

 

30.00

 

 

 

 

 

 

mmHg

 

hPa

 

inH

2

0

 

 

1560.0

 

2070.0

 

800.0

 

 

 

 

 

 

SPECIFICATION 

 
Model: KANE3500-30 
 

Nominal Pressure Ranges 

 

mBar

 

kPa

 

PSI

 

 

2070.0

 

207.00

 

30.00

 

 

 

 

 

 

mmHg

 

hPa

 

inH

2

0

 

 

1560.0

 

2070.0

 

800.0

 

 

 

 

Содержание 18853

Страница 1: ...KANE3500 PRESSURE METER Stock No 18853 January 2010 KANE3500 PRESSURE METER Stock No 18853 January 2010...

Страница 2: ...can be live data frozen data or stored data 255 sets of tests can be stored in non volatile memory GENERAL OPERATION When the instrument is first switched on using the top line of the display shows i...

Страница 3: ...The meter counts down from 10 after OFF is pressed Press to cancel OFF MENU Press and hold for MENU Two lines of 20 characters can be added to the header of printouts The meters are controlled using 8...

Страница 4: ...KLIGHT HOLD PRINT A quick press to HOLD the reading and the display flashes A longer press to PRINT ZERO Press and hold until there is a beep to ZERO the reading BACKLIGHT Press and hold until there i...

Страница 5: ...arrows also change settings such as date time and other menu items when MENU mode has been selected UP DOWN Press either to change the bottom line of the display ENTER Press to cancel OFF The buttons...

Страница 6: ...ry Set the meter s correct time date and units These are then stored when the meter is switched off 1 BEFORE USING THE METER FOR THE FIRST TIME Remove the cover and fit a new battery in the battery co...

Страница 7: ...measure Excessive pressure 5 times nominal range can permanently damage the meter s pressure sensor 2 BEFORE USING THE METER EVERY TIME After switch on check that date and time are correct and battery...

Страница 8: ...input measurements of pressure or vacuum Only use the input when taking a differential measurement NOTE Use the input for all single input measurements of pressure or vacuum Only use the input when t...

Страница 9: ...ss and use the button to select the following function for change STORE SETUP PRESSURE SERVICE EXIT 3 THE MENU FUNCTIONS The MENU structure comprises main menu options STORE SETUP and PRESSURE have a...

Страница 10: ...e key and change using the keys the function The final logical returns you to the main menu display Press to select a main menu function When you have selected the function to change press to select R...

Страница 11: ...nction the final logical is pressed No changes are made if the MENU function is aborted by pressing To exit the MENU function the final logical is pressed No changes are made if the MENU function is a...

Страница 12: ...f decimal places LOW one less decimal place displayed PS UNITS mBAR inH2O hPa mmHg PSI kPa Pa mmH2O PRESSURE SMOOTH OFF Normal ON Readings filtered to damp out fluctuations RESOLVE HIGH maximum number...

Страница 13: ...me 2 Stabilisation Period Time 3 Tightness Test duration Set using Times are in minutes 01 1 Minute REPORT TEST VIEW DEL ALL SETTING NB Automated Let By Test Stabilisation Tightness Test Display store...

Страница 14: ...user defined format Header 2 lines each of up to 20 characters that appear on the printout Use and SETUP 1 2 3 4 5 Temperature units o F or o C AUTO OFF adjusts the auto power off time A setting of 0...

Страница 15: ...6 7 8 Language English French Italian German Spanish The display s contrast EXIT 6 7 8 Language English French Italian German Spanish The display s contrast EXIT...

Страница 16: ...o exit this mode LOG Saves a set of results when is pressed Storage capacity is 255 sets of results STORE VIEW The test number appears on the top line Use or to change test number The bottom line show...

Страница 17: ...seconds and then press to activate and exit DEL ALL Clears all the memory when YES and pressed EXIT AUTO STO Automatically stores readings at a preset interval Press then select YES or NO If YES set t...

Страница 18: ...OPTIONS Max Pressure Min Pressure Time Date Pressure units Battery Status Instrument Internal Temp BOTTOM LINE OPTIONS Max Pressure Min Pressure Time Date Pressure units Battery Status Instrument Int...

Страница 19: ...Max Pressure is displayed as P 1 3 1 9 BOTTOM LINE OPTIONS Min Pressure is displayed as P 1 3 _ 1 1 Max Pressure is displayed as P 1 3 1 9 BOTTOM LINE OPTIONS Min Pressure is displayed as P 1 3 _ 1 1...

Страница 20: ...zero the meter in the position you are taking the readings and hold the meter still during the test 4 MEASURING Make sure you do not exceed the meter s operating specifications Do not exceed the meter...

Страница 21: ...e channel measurements pressure or vacuum Only use the left hand port for differential measurements Differential Port Port 1 IR Printer Output Always use the top right hand port Port 1 for taking sing...

Страница 22: ...switched on for at least five minutes and then re zeroed before taking such sensitive measurements If the pressure being measured exceeds the meter s design range the display will show OR for over ra...

Страница 23: ...PRINTOUTS The standard printout is The MAX and MIN readings are those observed since last switch on EXAMPLE PRINTOUTS The standard printout is The MAX and MIN readings are those observed since last sw...

Страница 24: ...NOTE Printouts of stored readings will also include the TEST NO below the header NOTE Printouts of stored readings will also include the TEST NO below the header...

Страница 25: ...n minutes TIME 01 Let By Test duration TIME 02 Stabilisation period TIME 03 Tightness Test duration 5 LET BY and TIGHTNESS TESTING only applicable for KANE3500 1 UK and KANE3500 2 UK To set the durati...

Страница 26: ...d adjust the pressure accordingly Once the correct pressure has been achieved press TIME 59 P2 10 05 LET BY TEST only applicable for KANE3500 1 UK and KANE3500 2 UK Press select PRESSURE select REPORT...

Страница 27: ...it the test sequence If the system passes the let by test press to continue with the tightness test The display shows the countdown timer Time 1 in REPORT SETTINGS and the measured pressure Once the c...

Страница 28: ...adjust the pressure accordingly Once the correct pressure has been achieved press TIME 59 P3 19 95 TIGHTNESS TEST only applicable for KANE3500 1 UK and KANE3500 2 UK After completing the let by test...

Страница 29: ...shed P3 is stored and the display changes to TIME 120 P4 19 85 Where TIME Time 3 in REPORT SETTINGS The display shows the countdown timer Time 2 in REPORT SETTINGS and the measured pressure Once the c...

Страница 30: ...ssure at the end of the test period Press to log the result The display shows the log number LOG 03 At the end of the countdown period the display automatically changes to P3 19 87 P4 19 84 Where P3 i...

Страница 31: ...o print if required The logged data can always be accessed by VIEW in the report menu Press to EXIT Press PRINT to print if required The logged data can always be accessed by VIEW in the report menu P...

Страница 32: ...tness Test Printout NOTE Printouts of stored readings will also include the LOG NO Below the header Let By Test and Tightness Test Printout NOTE Printouts of stored readings will also include the LOG...

Страница 33: ...ion is performed in firmware and there are no user accessible adjustments or user serviceable parts 6 METER ANNUAL RECALIBRATION AND SERVICE The meter should be re calibrated and serviced annually by...

Страница 34: ......

Страница 35: ...Pressure Ranges mBar mmH2O Pa kPa 80 00 800 0 8000 8 000 PSI mmHg hPa inH20 1 000 60 00 80 00 30 00 7 SPECIFICATION Model KANE3500 1 Nominal Pressure Ranges mBar mmH2O Pa kPa 80 00 800 0 8000 8 000 PS...

Страница 36: ...re Ranges mBar mmH2O Pa kPa 160 00 1600 0 16000 16 000 PSI mmHg hPa inH20 2 000 120 00 160 00 60 00 SPECIFICATION Model KANE3500 2 Nominal Pressure Ranges mBar mmH2O Pa kPa 160 00 1600 0 16000 16 000...

Страница 37: ...Pressure Ranges mBar mmH2O Pa kPa 400 0 4000 32000 40 00 PSI mmHg hPa inH20 5 000 300 0 400 0 150 0 SPECIFICATION Model KANE3500 5 Nominal Pressure Ranges mBar mmH2O Pa kPa 400 0 4000 32000 40 00 PSI...

Страница 38: ...E3500 15 Nominal Pressure Ranges mBar kPa PSI 1040 0 100 00 15 00 mmHg hPa inH20 780 0 1040 0 400 0 SPECIFICATION Model KANE3500 15 Nominal Pressure Ranges mBar kPa PSI 1040 0 100 00 15 00 mmHg hPa in...

Страница 39: ...3500 30 Nominal Pressure Ranges mBar kPa PSI 2070 0 207 00 30 00 mmHg hPa inH20 1560 0 2070 0 800 0 SPECIFICATION Model KANE3500 30 Nominal Pressure Ranges mBar kPa PSI 2070 0 207 00 30 00 mmHg hPa in...

Страница 40: ...uracy 20 Pa 0 1 Pa 0 5 Pa 100 Pa 0 1 Pa 3 Pa 2000 Pa 1 Pa 3 of reading 80 hPa 0 01 hPa 3 of reading Pressure Measurement Model KANE3500 1 Max overrange without damage to sensor is 5 times nominal rang...

Страница 41: ...range Range Resolution Accuracy 20 Pa 0 3 Pa 2 Pa 2000 Pa 0 3 Pa 5 Pa 160 hPa 0 01 hPa 3 of reading Pressure Measurement Model KANE3500 2 Max overrange without damage to sensor is 5 times nominal rang...

Страница 42: ...Range Resolution Accuracy 32000 Pa 1 Pa 5 Pa 300 hPa 0 01 hPa 0 05 hPa 400 hPa 0 1 hPa 3 of reading Pressure Measurement Model KANE3500 5 Max overrange without damage to sensor is 5 times nominal rang...

Страница 43: ...nominal range Range Resolution Accuracy 309 99 hPA 0 01 hPa 10 hPa 1040 0 hPA 0 1 hPa 3 of reading Pressure Measurement Model KANE3500 15 Max overrange without damage to sensor is 5 times nominal ran...

Страница 44: ...times nominal range Range Resolution Accuracy 1076 hPA 0 1 hPa 10 hPa 2070 0 hPA 1 hPa 3 of reading Pressure Measurement Model KANE3500 30 Max overrange without damage to sensor is 5 times nominal ran...

Страница 45: ...x 80mm x 40mm 180mm incl spigots 0 C to 45 C 10 to 90 RH non condensing 9 Volt PP3 alkaline battery Dimensions Weight Handset Ambient Operating Range Power Supply 295 grams with battery 160mm x 80mm x...

Страница 46: ...romagnetic radiation levels and occasionally it may be appropriate to check the meter before use by 8 ELECTROMAGNETIC COMPATIBILITY European Council Directive 89 336 EEC requires electronic equipment...

Страница 47: ...osition to minimize interference or switch off if possible the offending equipment during your test Use the normal start up sequence in the location where the meter will be used Switch on all localize...

Страница 48: ...nterference has occurred and this advice is only given to satisfy the requirements of the Directive At the time of writing this manual Jan 2010 Kane International Ltd are not aware of any field based...

Страница 49: ...cation EC EMC KI KANE3500 details the specific test configuration performance and conditions of use ELECTROMAGNETIC COMPATIBILITY This product has been tested for compliance with the following generic...

Страница 50: ...ent to the appropriate recycling centre in accordance with current legislation and local guidelines Please Note Batteries used in this instrument should be disposed of in accordance with current legis...

Страница 51: ...nd year The order can be changed using the menu function Date is recorded when each test is printed APPENDIX 1 Main Parameters UNITS DATE User selectable from PSI hPa inH2O mBar mmH2O Pa kPa NOTE Not...

Страница 52: ...date and time for up to one minute if outside this time it may be necessary to re enter the details TIME The time is shown in hours and minutes expressed in Military time or the 24 hr clock Time is re...

Страница 53: ...than 10 of charge and should be replaced readings may be affected if used with low power batteries SYMBOLS used on the display T P The internal temperature of the meter The pressure reading in the us...

Страница 54: ......

Страница 55: ......

Страница 56: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Отзывы: